     * Additional information about each of the calculations is provided in the [[test-and-measurement/current-and-power-adapter/reference-workspace|]].     * Additional information about each of the calculations is provided in the [[test-and-measurement/current-and-power-adapter/reference-workspace|]].
   - Save the modified workspace with altered name (to keep the original unchanged).    - Save the modified workspace with altered name (to keep the original unchanged). 
 +
 +
 +Common issues with the workspace and how to resolve them: 
 +  * In the Measurements window, the Reactive Power (Q) and Phase shift of Current relative to Voltage (variant Phi) list a "Syntax Error"
 +    * This is likely occurring because you are using shift or scan mode at a low sample rate. The two calculations in question are based the values between the time cursors. Unlike Repeated mode, where the whole buffer of data is collected before being rendered and then calculated upon, Scan and Shift modes fill up their data over time, resulting in a situation where WaveForms is trying to calculate on data that hasn't come in yet. Once the Scope Plot is filled with data, simply right click on the Measurement line item with the Syntax Error, choose Edit, then click okay without changing anything at all to resolve the issue.
 +  * My current and power in the plot seem to be flat instead of fluctuating like I would expect with my variable power input. 
 +    * Make sure the positive 5 V supply is enabled.
 +  * I clicked run all but nothing is happening
 +    * Check to see if the Scope instrument started running. 
 +    * Confirm that your load is ready to receive power and isn't switched off.
 +    * If you adjusted the timebase to be much wider, change the Scope mode from Repeated to Scan or Shift so that WaveForms isn't collecting all 5 or 10 seconds worth of data before displaying it on the screen.
  
 ---- ----