Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
test-and-measurement:analog-discovery-pro-5250:reference-manual [2021/10/11 17:02] – fixed some image links James Colvintest-and-measurement:analog-discovery-pro-5250:reference-manual [2022/10/05 15:57] (current) – changed 'Scopes and Instruments' to 'Test and Measurement' Fausto Soares
Line 1: Line 1:
-====== Analog Discovery Pro 5250 Reference Manual ======+====== Analog Discovery Pro (ADP5250) Reference Manual ====== 
 ~~TechArticle~~ ~~TechArticle~~
 /*| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:analog-discovery-pro-3450-front-1000x612.png?direct |}}*/ /*| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:analog-discovery-pro-3450-front-1000x612.png?direct |}}*/
Line 6: Line 7:
  
 | image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250-obl.png?direct|}} | image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250-obl.png?direct|}}
-| image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250.png?direct|}} +| image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250-front.png?direct|}} 
-| image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250rear.png?direct|}}+| image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250-rear.png?direct|}}
 | image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250kit.jpg?direct|}} | image = {{ :test-and-measurement:analog-discovery-pro-5250:adp5250kit.jpg?direct|}}
-| image = {{ :test-and-measurement:analog-discovery-pro-5250:oscilloscope.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:waveform-generator.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:power-supplies.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:logic-analyzer.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:static-io.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:spectrum-analyzer.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:network-analyzer.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:impedance-analyzer.png?direct |}} 
-| image = {{ :reference:test-and-measurement:analog-discovery-pro-3x50:script-editor.png?direct |}} 
-| image = {{ :test-and-measurement:analog-discovery-pro-5250:connector.png?direct |}} 
-}} 
- 
  
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-scope.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-wavegen.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-supplies.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-dmm-run.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-logger.png|}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-logic.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-staticio.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-spectrum.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-network.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-impedance.png}}
 +| image = {{:test-and-measurement:analog-discovery-pro-5250:instrument-script.png|}}
 +}}
  
 ===== Oscilloscope ===== ===== Oscilloscope =====
  
-{{reference:test-and-measurement:analog-discovery-pro-5250:oscilloscope.png?nolink&800}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-scope.png?nolink&800}}
  
-The Analog Discovery Pro ADP5250 can be used with WaveForms’ Oscilloscope instrument to capture analog data via the 2 analog input (“Scope”) channels using BNC probes. +The Analog Discovery Pro (ADP5250can be used with WaveForms’ Oscilloscope instrument to capture analog data via the 2 analog input  channels using BNC probes. 
  
 Since the Analog Discovery Pro’s analog input channels are shared, the Oscilloscope instrument cannot be used at the same time as the Spectrum Analyzer, Network Analyzer, or Impedance Analyzer instruments. Since the Analog Discovery Pro’s analog input channels are shared, the Oscilloscope instrument cannot be used at the same time as the Spectrum Analyzer, Network Analyzer, or Impedance Analyzer instruments.
 +
 +The Analog Discovery Pro features an additional connector, found just to the right of the analog input channels, that provides a square wave for use in compensating your probes. By connecting the probe to the tab indicated by the square wave symbol and the ground clip to the tab indicated by the ground symbol, a 5 V, 1 kHz, square wave reference signal can be measured using the Scope instrument. If your probes support adjusting the compensation, you can then do so, by minimizing the over/undershoot of the measured reference signal.
  
 For more information on the analog input ("Scope") channels, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Oscilloscope instrument, please visit the [[test-and-measurement:guides:waveforms-oscilloscope]] guide. For more information on the analog input ("Scope") channels, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Oscilloscope instrument, please visit the [[test-and-measurement:guides:waveforms-oscilloscope]] guide.
Line 43: Line 46:
   * Captured data files can be exported in standard formats   * Captured data files can be exported in standard formats
   * Scope configurations can be saved, exported, and imported   * Scope configurations can be saved, exported, and imported
 +  * Front panel connector generating a 5 V, 1 kHz, square wave for convenient compensation of BNC oscilloscope probes
  
 ---- ----
 ===== Waveform Generator ===== ===== Waveform Generator =====
  
-{{reference:test-and-measurement:analog-discovery-pro-5250:waveform-generator.png?nolink&800}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-wavegen.png?nolink&800}}
  
-The Analog Discovery Pro ADP5250 can be used with WaveForms' Wavegen instrument to output analog voltage waves via BNC cable. The Waveform Generator converts 14-bit digital samples to analog at rate of up to 125 MS/s on a single channel. When the Wavgen instrument is used, the Analog Discovery Pro's analog output channel act as an Arbitrary Waveform Generator. The instrument supports everything from simple waveforms like Sine and Triangle waves or custom sets of samples can be defined by the user in applications like Excel and imported to WaveForms.+The Analog Discovery Pro (ADP5250can be used with WaveForms' Wavegen instrument to output analog voltage waves via BNC cable. The Waveform Generator converts digital samples to analog and stimulates circuit connected over a single channel. When the Wavegen instrument is used, the Analog Discovery Pro's analog output channel acts as an Arbitrary Waveform Generator. The instrument supports everything from simple waveforms like Sine and Triangle waves or custom sets of samples can be defined by the user in applications like Excel and imported to WaveForms.
  
-The waveform generator channel is considered a single ended pin, however, a connected circuit must share a ground with the Analog Discovery Pro. The channel has a bandwidth of >15MHz through the BNC connector. AC amplitudes of +-6V and DC offsets of +-6V are supported.+The waveform generator channel is considered a single ended pin, however, a connected circuit must share a ground with the Analog Discovery Pro.
  
 Since the Analog Discovery Pro's analog output channel is shared, the Waveform Generator instrument cannot be used at the same time as the Network Analyzer, or Impedance Analyzer instruments. Since the Analog Discovery Pro's analog output channel is shared, the Waveform Generator instrument cannot be used at the same time as the Network Analyzer, or Impedance Analyzer instruments.
Line 64: Line 68:
 ===== Power Supplies ===== ===== Power Supplies =====
  
-{{:test-and-measurement:analog-discovery-pro-5250:supplies.png?600|}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-supplies.png?nolink&800}}
  
-The Analog Discovery Pro 5250 has 3 programmable Digital Voltage Supplies, which can supply between 0V to +6V0V to +25V, and 0V to -25V through the Power Supplies ("Supplies") instrument.+The Analog Discovery Pro (ADP5250) has 3 programmable Digital Voltage Supplies, which can supply between 0 V to +6 V0 V to +25 V, and 0 V to -25 V through the Power Supplies ("Supplies") instrument.
  
-For more information on using the programmable power supplies, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Power Supplies instrument, please visit the [[test-and-measurement:guides:waveforms-supplies]] guide.+Each of the power supply rails is associated with an LED available on the connector, just above the supply's screw terminal, which is used to indicate the operating mode of that supply. When a load doesn't require more current than the supply's current limit, the supply operates in Constant Voltage (CV) mode, and the indicator LED is blue. When the load does try to draw more current than the current limit, the supply operates in Constant Current (CC) mode, and the LED is orange. WaveForms' Supplies instrument UI for the ADP5250 also features these LEDs, showing CV mode with a green indicator and CC with a red indicator. 
 + 
 +For more information on using the programmable power supplies, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Power Supplies instrument, please visit the [[test-and-measurement:guides:waveforms-readback-supplies]] guide.
  
 ==== Features ==== ==== Features ====
   * Three programmable power supplies   * Three programmable power supplies
-    * 0V to 6V (0 A to 1 A) +    * 0 V to 6 V (0 A to 1 A) 
-    * 0V to +25V (0 mA to 500 mA) +    * 0 V to +25 V (0 mA to 500 mA) 
-    * 0V to -25V (0 mA to 500 mA)+    * 0 V to -25 V (0 mA to 500 mA)
   * Up to 1A output current   * Up to 1A output current
 +
 +----
 +===== Data Logger =====
 +
 +{{:test-and-measurement:analog-discovery-pro-5250:instrument-logger.png?nolink&800|}}
 +
 +The Analog Discovery Pro (ADP5250) can be used with WaveForms' "Logger" instrument in order to capture large buffers of analog input data on both Scope channels.
 +
 +The Data Logger can capture buffers of data at update rates of up to 10 samples per second. The maximum duration of a log is dependent on the update rate, but at the extreme, can run for over a thousand hours.
 +
 +Since the Analog Discovery Pro's analog input channels are shared, the Data Logger instrument cannot be used at the same time as the Oscilloscope, Spectrum Analyzer, Network Analyzer, or Impedance Analyzer instruments.
 +
 +For more information on the analog input ("Scope") channels, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Data Logger instrument, please visit the [[test-and-measurement:guides:waveforms-data-logger]] guide.
 +
 +==== Features ====
 +  * Measurements: DC, AC RMS, True RMS, with Averages, Minimums, and Maximums
 +  * Up to 24 hours of data logged at a 1 Hz sample rate
 +  * Scriptable conversion functions
  
 ---- ----
 ===== Logic Analyzer ===== ===== Logic Analyzer =====
  
-{{reference:test-and-measurement:analog-discovery-pro-5250:logic-analyzer.png?nolink&800}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-logic.png?nolink&800}}
  
-The Analog Discovery Pro 5250 can be used with WaveForms' "Logic" instrument to act as a Logic Analyzer. When used this way, the 34 digital input/output channels are configured to capture high/low logic states on connected pins at a sample rate of up to 100 MS/s. These channels are tolerant of voltages of up to 5V when configured as inputs.+The Analog Discovery Pro (ADP5250) can be used with WaveForms' "Logic" instrument to act as a Logic Analyzer. When used this way, the digital input channels (in the "Logic Analyzer" connector, part of the Mixed Signal Oscilloscope) are configured to capture the high/low logic states over time of connected digital signals.
  
 Individual input/output channels can be grouped as buses and protocols. Protocol groups can be used to view the decoded contents of packets of many common communications protocols, including SPI, I2C, UART, CAN, and I2S. Individual input/output channels can be grouped as buses and protocols. Protocol groups can be used to view the decoded contents of packets of many common communications protocols, including SPI, I2C, UART, CAN, and I2S.
Line 102: Line 126:
 ===== Static I/O ===== ===== Static I/O =====
  
-{{reference:test-and-measurement:analog-discovery-pro-5250:static-io.png?nolink&800}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-staticio.png?nolink&800}}
  
-The Analog Discovery Pro 5250 can be used with WaveForms' Static I/O instrument to emulate a variety of user input/output devices on the digital input/output pins. Virtual LEDs, buttons, switches, sliders, and displays can be assigned to specific digital I/O pins, and interacted with within the WaveForms user interface.+The Analog Discovery Pro (ADP5250) can be used with WaveForms' Static I/O instrument to emulate a variety of user input/output devices on the digital input/output pins. Virtual LEDs, buttons, switches, sliders, and displays can be assigned to specific digital I/O pins, found on the "Digital I/O" connector, and interacted with within the WaveForms user interface.
  
-**Important Note:** //To prevent damage to the device, care must be taken not to drive input signals to the digital input/output channels over 5V.//+**Important Note:** //To prevent damage to the device, care must be taken not to drive input signals to the digital input/output channels over 5 V.//
  
 Digital input/output channels used by the Static I/O instrument can still be used by other instruments using the same digital input/output channels, however, other instruments can only use these shared channels as inputs. Digital input/output channels used by the Static I/O instrument can still be used by other instruments using the same digital input/output channels, however, other instruments can only use these shared channels as inputs.
Line 119: Line 143:
 ===== Spectrum Analyzer ===== ===== Spectrum Analyzer =====
  
-{{reference:test-and-measurement:analog-discovery-pro-5250:spectrum-analyzer.png?nolink&800}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-spectrum.png?nolink&800}}
  
-The Analog Discovery Pro 5250 can be used with WaveForms' "Spectrum" instrument to view the power of frequency-domain components of analog signals captured on the analog input channels.+The Analog Discovery Pro (ADP5250) can be used with WaveForms' "Spectrum" instrument to view the power of frequency-domain components of analog signals captured on the analog input channels.
  
-Signals with minimum/maximum frequencies between 0 Hz and 50Mhz can be plotted in units of peak voltage, RMS voltage, and various voltage level ratio units.+Signals can be plotted in units of peak voltage, RMS voltage, and various voltage level ratio units.
  
 Since the Spectrum Analyzer instrument uses the same hardware resources as the Oscilloscope, Network Analyzer, and Impedance Analyzer instruments, it cannot be used at the same time as these other instruments. Since the Spectrum Analyzer instrument uses the same hardware resources as the Oscilloscope, Network Analyzer, and Impedance Analyzer instruments, it cannot be used at the same time as these other instruments.
Line 140: Line 164:
 ---- ----
 ===== Network Analyzer ===== ===== Network Analyzer =====
-{{reference:test-and-measurement:analog-discovery-pro-5250:network-analyzer.png?nolink&800}} 
  
-The Analog Discovery Pro 5250 can be used with WaveForms' "Network" instrument to view the amplitude and phase response of a circuit under test. Nichols and Nyquist plots can also be viewed with this instrument.+{{:test-and-measurement:analog-discovery-pro-5250:instrument-network.png?nolink&800}} 
 + 
 +The Analog Discovery Pro (ADP5250) can be used with WaveForms' "Network" instrument to view the amplitude and phase response of a circuit under test. Nichols and Nyquist plots can also be viewed with this instrument.
  
 Frequency sweeps can be performed in ranges between 1 mHz and 10 MHz with up to 10k samples per decade. The wave used for the sweep can be customized, and uses the same resources as the Waveform Generator instrument. Frequency sweeps can be performed in ranges between 1 mHz and 10 MHz with up to 10k samples per decade. The wave used for the sweep can be customized, and uses the same resources as the Waveform Generator instrument.
  
-The Network Analyzer instrument uses the analog output and analog input channels of the Analog Discovery Pro 5250 to probe a test circuit. The Network Analyzer can be configured to use an external signal to provide input to the circuit under test, rather than using the analog output channels.+The Network Analyzer instrument uses the analog output and analog input channels of the Analog Discovery Pro (ADP5250) to probe a test circuit. The Network Analyzer can be configured to use an external signal to provide input to the circuit under test, rather than using the analog output channels.
  
-Since the Analog Discovery Pro's analog input and output channels are shared, the Network Analyzer instrument cannot be used at the same time as the Oscilloscope, Waveform Generator, Voltmeter, Data Logger, Spectrum Analyzer, or Impedance Analyzer instruments.+Since the Analog Discovery Pro's analog input and output channels are shared, the Network Analyzer instrument cannot be used at the same time as the Oscilloscope, Waveform Generator, Data Logger, Spectrum Analyzer, or Impedance Analyzer instruments.
  
 For more information on the analog output and analog input channels, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Network Analyzer instrument, please visit the [[test-and-measurement:guides:waveforms-network-analyzer]] guide. For more information on the analog output and analog input channels, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Network Analyzer instrument, please visit the [[test-and-measurement:guides:waveforms-network-analyzer]] guide.
Line 160: Line 185:
 ===== Impedance Analyzer ===== ===== Impedance Analyzer =====
  
-{{reference:test-and-measurement:analog-discovery-pro-5250:impedance-analyzer.png?nolink&800}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-impedance.png?nolink&800}}
  
-The Analog Discovery Pro 5250 can be used with WaveForms' "Impedance" instrument to view a wide variety of frequency response characteristics of a circuit under test. Input, Phase, Voltage, Current, Impedance, Admittance, Inductance, Factor, and Nyquist plots are all available. In addition, Custom plots can be used to present the results of a wide variety of different mathematical operations on buffered data.+The Analog Discovery Pro (ADP5250) can be used with WaveForms' "Impedance" instrument to view a wide variety of frequency response characteristics of a circuit under test. Input, Phase, Voltage, Current, Impedance, Admittance, Inductance, Factor, and Nyquist plots are all available. In addition, Custom plots can be used to present the results of a wide variety of different mathematical operations on buffered data.
  
-Frequency sweeps can be performed with in ranges between 100 uHz and 25 MHz, with as many as 10k samples per decade. The signal used to perform the sweep can be selected from a variety of preset, with configurable amplitude and offset. An external network analyzer reference circuit can be selected from a variety of options.+Frequency sweeps are performed to stimulate the connected circuit and measure the response. The signal used in the sweeps can be selected from a variety of presets, with configurable amplitude and offset. An external network analyzer reference circuit can be selected from a variety of options.
  
 The Impedance Analyzer instrument uses the Analog Discovery Pro's analog output channels and analog input channels to probe a test circuit. The Impedance Analyzer instrument uses the Analog Discovery Pro's analog output channels and analog input channels to probe a test circuit.
  
-Since the Analog Discovery Pro's analog input and output channels are shared, the Impedance Analyzer instrument cannot be used at the same time as the Oscilloscope, Waveform Generator, Voltmeter, Data Logger, Spectrum Analyzer, or Network Analyzer instruments.+Since the Analog Discovery Pro's analog input and output channels are shared, the Impedance Analyzer instrument cannot be used at the same time as the Oscilloscope, Waveform Generator, Data Logger, Spectrum Analyzer, or Network Analyzer instruments.
  
 For more information on the analog output and analog input channels, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Impedance Analyzer instrument, please visit the [[test-and-measurement:guides:waveforms-impedance-analyzer]] guide. For more information on the analog output and analog input channels, please visit the [[test-and-measurement:analog-discovery-pro-5250:specifications]]. For a walkthrough of the different features of WaveForms' Impedance Analyzer instrument, please visit the [[test-and-measurement:guides:waveforms-impedance-analyzer]] guide.
Line 180: Line 205:
 ---- ----
 ===== WaveForms Script Editor ===== ===== WaveForms Script Editor =====
-{{reference:test-and-measurement:analog-discovery-pro-5250:script-editor.png?nolink&800}}+{{:test-and-measurement:analog-discovery-pro-5250:instrument-script.png?nolink&800|}}
  
 Each of WaveForms' instruments can be controlled through scripts within the WaveForms application itself. WaveForms' "Script" instrument allows the user to write and run JavaScript code that can control the rest of the application through an extensive API. This allows the user to configure and run many instruments at the same time, in an easily repeatable way. Each of WaveForms' instruments can be controlled through scripts within the WaveForms application itself. WaveForms' "Script" instrument allows the user to write and run JavaScript code that can control the rest of the application through an extensive API. This allows the user to configure and run many instruments at the same time, in an easily repeatable way.
  
-A variety of code examples are available in the application to aid in learning to write WaveForms scripts. Additional resources for writing scripts can be found on the Scopes and Instruments section of the [[https://forum.digilentinc.com|Digilent Forum]].+A variety of code examples are available in the application to aid in learning to write WaveForms scripts. Additional resources for writing scripts can be found on the Test and Measurement section of the [[https://forum.digilent.com|Digilent Forum]].
  
 A plot pane within the Script instrument itself can be used to integrate data from many different instruments, and display it in a highly customizable way. A plot pane within the Script instrument itself can be used to integrate data from many different instruments, and display it in a highly customizable way.
Line 206: Line 231:
   * Provides control of hardware channels and virtual instruments to custom applications   * Provides control of hardware channels and virtual instruments to custom applications
  
-**Note:** For custom application we recommend and support WaveForms SDK, however, NI’s VirtualBench C API and VirtualBench LabVIEW Vis can also be used.+**Note:** For custom applications we recommend and support WaveForms SDK, however, NI’s VirtualBench C API and VirtualBench LabVIEW Vis can also be used.
  
 ---- ----