===== Introduction ===== | ===== Introduction ===== | ||
Line 10: | Line 10: | ||
===== Prerequisites ===== | ===== Prerequisites ===== | ||
* A Digilent Test & Measurement Device with Analog Input and Output Channels: | * A Digilent Test & Measurement Device with Analog Input and Output Channels: | ||
- | * [[test-and-measurement: | + | |
+ | | ||
* [[test-and-measurement: | * [[test-and-measurement: | ||
+ | * [[test-and-measurement: | ||
+ | * [[test-and-measurement: | ||
* [[test-and-measurement: | * [[test-and-measurement: | ||
* [[test-and-measurement: | * [[test-and-measurement: | ||
Line 19: | Line 22: | ||
---- | ---- | ||
===== Guide ===== | ===== Guide ===== | ||
- | ==== 1. Opening the WaveGen | + | ==== 1. Opening the Wavegen |
<WRAP group> | <WRAP group> | ||
Line 26: | Line 29: | ||
Plug in the Test & Measurement Device, then start WaveForms and make sure the device is connected. | Plug in the Test & Measurement Device, then start WaveForms and make sure the device is connected. | ||
- | If no device is connected to the host computer when WaveForms launches, the **Device Manager** will be launched. Make sure that the device is plugged in and turned on, at which point it will appear in the Device Manager' | + | If no device is connected to the host computer when WaveForms launches, the **Device Manager** will be launched. Make sure that the device is plugged in and turned on, at which point it will appear in the Device Manager' |
**Note:** //" | **Note:** //" | ||
- | **Note:** //The Device Manager can be opened by clicking on the " | + | **Note:** //The Device Manager can be opened by clicking on the " |
</ | </ | ||
<WRAP column half> | <WRAP column half> | ||
- | {{ reference: | + | {{: |
</ | </ | ||
</ | </ | ||
Line 41: | Line 44: | ||
Once the Welcome page loads, in the instrument panel at the left side of the window, click on the **Wavegen** button to open the //Waveform Generator// instrument. | Once the Welcome page loads, in the instrument panel at the left side of the window, click on the **Wavegen** button to open the //Waveform Generator// instrument. | ||
</ | </ | ||
- | {{ reference: | + | {{: |
</ | </ | ||
---- | ---- | ||
Line 53: | Line 56: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{learn:instrumentation: | + | {{:test-and-measurement: |
</ | </ | ||
</ | </ | ||
- | ==== 2. Using the WaveGen | + | ==== 2. Using the Wavegen |
//This section details how to configure and run the Wavegen instrument// | //This section details how to configure and run the Wavegen instrument// | ||
Line 75: | Line 78: | ||
<-- | <-- | ||
---- | ---- | ||
- | == BNC Cables | + | == BNC Oscilloscope Probes |
- | //If your Digilent Test and Measurement device supports BNC cables | + | //If your Digilent Test and Measurement device supports BNC cables, |
+ | |||
+ | --> Analog Discovery 3 with BNC Adapter # | ||
+ | <WRAP group> <WRAP column half> | ||
+ | Connect the [[test-and-measurement: | ||
+ | |||
+ | **Note:** //For this demo, both ground wires were not connected together, because the Oscilloscope channel shares the same ground as the Wavegen channel. While BNC Probes are single-ended (as is the Waveform Generator hardware), a connected circuit must still share a common ground with the device. No more than one ground should be connected, in order to avoid the creation of ground loops, which may damage your device.// | ||
+ | |||
+ | **Note:** //When using BNC Probes, make sure to take note of the probes' | ||
+ | </ | ||
+ | {{: | ||
+ | </ | ||
+ | |||
+ | <WRAP group> <WRAP column half> | ||
+ | Take note of the jumper that selects between AC and DC coupling on the BNC adapter. Either will work to measure the sine wave produced here, but some waveforms may find one setting or the other significantly more useful. | ||
+ | </ | ||
+ | {{: | ||
+ | </ | ||
+ | <-- | ||
--> Analog Discovery Studio # | --> Analog Discovery Studio # | ||
<WRAP group> <WRAP column half> | <WRAP group> <WRAP column half> | ||
- | Connect BNC cables to the Test and Measurement device' | + | Connect BNC cables to the Test and Measurement device' |
- | **Note:** //While BNC probes | + | **Note:** //For this demo, both ground wires were not connected together, because the Oscilloscope channel shares the same ground as the Wavegen channel. |
- | **Note:** //When using BNC cables, make sure to take note of the probes' | + | **Note:** //When using BNC Cables, make sure to take note of the probes' |
- | **Note:** //The Analog Discovery Studio' | + | **Note:** //The Analog Discovery Studio' |
**Note:** //The Analog Output signal produced by the Waveform Generator is available on both the BNC and MTE cables at the same time.// | **Note:** //The Analog Output signal produced by the Waveform Generator is available on both the BNC and MTE cables at the same time.// | ||
</ | </ | ||
- | {{: | + | {{: |
+ | </ | ||
+ | |||
+ | <WRAP group> <WRAP column half> | ||
+ | Make sure to flip the switch that selects between BNC and MTE input for Oscilloscope Channel 1 towards the BNC connector. | ||
+ | </ | ||
+ | {{: | ||
</ | </ | ||
<-- | <-- | ||
Line 97: | Line 124: | ||
--> Analog Discovery 2 with BNC Adapter # | --> Analog Discovery 2 with BNC Adapter # | ||
<WRAP group> <WRAP column half> | <WRAP group> <WRAP column half> | ||
- | Connect | + | Connect the [[test-and-measurement: |
- | **Note:** //While BNC Probes are single-ended (as is the Waveform Generator hardware), a connected circuit must still share a common ground with the device. No more than one ground should be connected, in order to avoid the creation of ground loops, which may damage your device.// | + | **Note:** //For this demo, both ground wires were not connected together, because the Oscilloscope channel shares the same ground as the Wavegen channel. |
- | **Note:** //When using BNC Cables, make sure to take note of their bandwidth. When these cables | + | **Note:** //When using BNC Probes, make sure to take note of the probes' |
</ | </ | ||
- | {{: | + | {{: |
</ | </ | ||
Line 109: | Line 136: | ||
Take note of the jumpers that select between 0 or 50 Ohm output impedance on the BNC adapter. Either setting will work for the loopback circuit used here, however, some applications may find one or the other more useful. | Take note of the jumpers that select between 0 or 50 Ohm output impedance on the BNC adapter. Either setting will work for the loopback circuit used here, however, some applications may find one or the other more useful. | ||
</ | </ | ||
- | {{: | + | {{: |
</ | </ | ||
<-- | <-- | ||
Line 117: | Line 144: | ||
Connect a BNC oscilloscope probe to the Analog Discovery Pro's Oscilloscope Channel 1 connector and a set of BNC minigrabbers to its Wavegen Out Channel 1 connector. Connect these together, as pictured, to form a loopback circuit. Check the input probe' | Connect a BNC oscilloscope probe to the Analog Discovery Pro's Oscilloscope Channel 1 connector and a set of BNC minigrabbers to its Wavegen Out Channel 1 connector. Connect these together, as pictured, to form a loopback circuit. Check the input probe' | ||
- | **Note:** //While BNC Probes are single-ended (as is the Waveform Generator hardware), a connected circuit must still share a common ground with the device. No more than one ground should be connected, in order to avoid the creation of ground loops, which may damage your device.// | + | **Note:** //For this demo, both ground wires were not connected together, because the Oscilloscope channel shares the same ground as the Wavegen channel. |
+ | |||
+ | **Note:** //When using BNC Probes, make sure to take note of the probes' | ||
</ | </ | ||
- | {{reference: | + | {{reference: |
</ | </ | ||
<-- | <-- | ||
Line 125: | Line 154: | ||
--> Analog Discovery Pro (ADP5250) # | --> Analog Discovery Pro (ADP5250) # | ||
<WRAP group> <WRAP column half> | <WRAP group> <WRAP column half> | ||
- | Connect a BNC oscilloscope probe to the Analog Discovery Pro's Oscilloscope Channel 1 connector and the scope probe or minigrabbers to the test signal and test signal ground pins on the front of the ADP5250 to create a loopback circuit. | + | Connect a BNC oscilloscope probe to the Analog Discovery Pro's Oscilloscope Channel 1 connector and the BNC minigrabbers to the AWG Channel 1 connector |
**Note:** //While BNC Probes are single-ended (as is the Waveform Generator hardware), a connected circuit must still share a common ground with the device. No more than one ground should be connected, in order to avoid the creation of ground loops, which may damage your device.// | **Note:** //While BNC Probes are single-ended (as is the Waveform Generator hardware), a connected circuit must still share a common ground with the device. No more than one ground should be connected, in order to avoid the creation of ground loops, which may damage your device.// | ||
Line 131: | Line 160: | ||
**Note:** //When using BNC Probes, make sure to take note of the probes' | **Note:** //When using BNC Probes, make sure to take note of the probes' | ||
</ | </ | ||
- | {{: | + | {{: |
</ | </ | ||
<-- | <-- | ||
Line 145: | Line 174: | ||
</ | </ | ||
- | {{ reference: | + | {{: |
</ | </ | ||
Line 158: | Line 187: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{ :learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 169: | Line 198: | ||
Regardless of the device, if BNC cables are used for input, make sure that the attenuation factor setting in WaveForms for Oscilloscope Channel 1 (pictured to the right) matches the setting of the probe. | Regardless of the device, if BNC cables are used for input, make sure that the attenuation factor setting in WaveForms for Oscilloscope Channel 1 (pictured to the right) matches the setting of the probe. | ||
</ | </ | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
<-- | <-- | ||
Line 184: | Line 213: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 200: | Line 229: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 216: | Line 245: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 227: | Line 256: | ||
=== 2.6 === | === 2.6 === | ||
- | Click the Run button in the control bar near the top of the window to begin output the signal. | + | Click the Run button |
</ | </ | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 244: | Line 273: | ||
=== 2.7 === | === 2.7 === | ||
- | Back in the Wavegen instrument, click Stop ({{:learn: | + | Back in the Wavegen instrument, click Stop ({{:test-and-measurement:analog-discovery-3: |
**Note:** //An idle channel will still output some amount of voltage, which is determined by the offset parameter by default. Unchecking enable at any point will stop the signal if it is running, disable the channel and set its output voltage to 0 V.// | **Note:** //An idle channel will still output some amount of voltage, which is determined by the offset parameter by default. Unchecking enable at any point will stop the signal if it is running, disable the channel and set its output voltage to 0 V.// | ||
Line 251: | Line 280: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 257: | Line 286: | ||
---- | ---- | ||
- | ==== Note: Analog Discovery Pro (ADP3450/ADP3250) Sample Rate ==== | + | //Digilent Oscilloscopes feature the ability to select from a range of sample rates, which affects many instruments. Click the dropdown below to learn how to use this feature.// |
- | {{section> | + | --> Modify the Digilent Oscilloscope System Clock Rate # |
+ | <WRAP group> <WRAP column half> | ||
+ | To access this setting, first open the Device Manager, which can be found in WaveForms' | ||
+ | The system clock rate itself can be changed by opening the Options dropdown, while the Digilent device is selected in the list of devices. By changing the selection in the " | ||
+ | |||
+ | **Note:** //Changing the system clock rate changes the set of sample rates that can be achieved by the Digilent Oscilloscope' | ||
+ | </ | ||
+ | {{ : | ||
+ | </ | ||
+ | <-- | ||
+ | |||
+ | ---- | ||
+ | ==== Note: Analog Discovery Pro (ADP5250) Wavegen Output Filter ==== | ||
+ | |||
+ | //The Analog Discovery Pro (ADP5250) features a switchable filter on the AWG output. Click the dropdown below to learn how to use this feature.// | ||
+ | |||
+ | --> Enable or Disable the ADP5250' | ||
+ | <WRAP group> <WRAP column half> | ||
+ | This filter can be enabled or disabled through the use of the dropdown on the right side of the control toolbar. It can be set to Auto, On, or Off. | ||
+ | </ | ||
+ | {{: | ||
+ | </ | ||
+ | <-- | ||
+ | |||
+ | ---- | ||
==== Note: Using Multiple Instruments ==== | ==== Note: Using Multiple Instruments ==== | ||
The //Waveform Generator// instrument can be used with other WaveForms instruments. Using multiple instruments at the same time is helpful in a wide variety of situations. One example would be to use the //Waveform Generator// to stimulate a circuit under test, the //Power Supplies// to provide power to the circuit and view the response with the // | The //Waveform Generator// instrument can be used with other WaveForms instruments. Using multiple instruments at the same time is helpful in a wide variety of situations. One example would be to use the //Waveform Generator// to stimulate a circuit under test, the //Power Supplies// to provide power to the circuit and view the response with the // | ||
Line 278: | Line 331: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 291: | Line 344: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 304: | Line 357: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 317: | Line 370: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 323: | Line 376: | ||
<WRAP group> <WRAP column half> | <WRAP group> <WRAP column half> | ||
- | The **Synchronization** dropdown allows the user to select how the Wavegen Channels operate in regards to each other. Selecting // | + | The **Synchronization** dropdown allows the user to select how the Wavegen Channels operate in regards to each other. Selecting // |
</ | </ | ||
- | {{ :learn: | + | {{:test-and-measurement:analog-discovery-3:waveforms-wavegen-triggers-sync.png?nolink&600 |}} |
</ | </ | ||
Line 331: | Line 384: | ||
**Edit** lets the user copy or swap output signal configurations between available channels. | **Edit** lets the user copy or swap output signal configurations between available channels. | ||
</ | </ | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
<WRAP group> | <WRAP group> | ||
<WRAP column half> | <WRAP column half> | ||
+ | |||
=== 3.2 The Preview Plot === | === 3.2 The Preview Plot === | ||
Line 344: | Line 398: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 360: | Line 414: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 375: | Line 429: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 388: | Line 442: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 403: | Line 457: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 417: | Line 471: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
</ | </ | ||
Line 430: | Line 484: | ||
<WRAP column half> | <WRAP column half> | ||
- | {{:learn: | + | {{:test-and-measurement: |
+ | |||
+ | </ | ||
+ | </ | ||
+ | |||
+ | /* | ||
+ | <WRAP group> | ||
+ | <WRAP column half> | ||
+ | |||
+ | FIXME The " | ||
+ | |||
+ | </ | ||
+ | <WRAP column half> | ||
+ | |||
+ | {{: | ||
</ | </ | ||
</ | </ | ||
+ | */ | ||
==== Next Steps ==== | ==== Next Steps ==== | ||
Line 443: | Line 512: | ||
For more information on WaveForms visit the [[software: | For more information on WaveForms visit the [[software: | ||
- | For technical support, please visit the [[https:// | + | For technical support, please visit the [[https:// |
{{tag> | {{tag> |