Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
test-and-measurement:analog-discovery-2:reference-manual [2022/02/18 23:05] – [Features] fix power supply features Arthur Brown | test-and-measurement:analog-discovery-2:reference-manual [2023/04/06 18:45] (current) – Arthur Brown | ||
---|---|---|---|
Line 39: | Line 39: | ||
The Analog Discovery 2 can be used with WaveForms' | The Analog Discovery 2 can be used with WaveForms' | ||
- | You can use the [[test-and-measurement: | + | You can use the [[test-and-measurement: |
- | When used with MTE cables, the oscilloscope channels' | + | When used with MTE cables, the oscilloscope channels' |
Since the Analog Discovery 2's analog input channels are shared, the Oscilloscope instrument cannot be used at the same time as the Voltmeter, Data Logger, Spectrum Analyzer, Network Analyzer, or Impedance Analyzer instruments. | Since the Analog Discovery 2's analog input channels are shared, the Oscilloscope instrument cannot be used at the same time as the Voltmeter, Data Logger, Spectrum Analyzer, Network Analyzer, or Impedance Analyzer instruments. | ||
Line 65: | Line 65: | ||
The Analog Discovery 2's GND reference is connected to the USB GND. Depending on the PC powering scheme, and other PC connections (Ethernet, audio, etc. – which might also be grounded) the Analog Discovery 2 GND reference might be connected to the whole GND system and ultimately to the power network protection (earth ground). The circuit under test might also be connected to earth or possibly floating. | The Analog Discovery 2's GND reference is connected to the USB GND. Depending on the PC powering scheme, and other PC connections (Ethernet, audio, etc. – which might also be grounded) the Analog Discovery 2 GND reference might be connected to the whole GND system and ultimately to the power network protection (earth ground). The circuit under test might also be connected to earth or possibly floating. | ||
- | For safety reasons, it is the user’s responsibility to understand the powering and grounding scheme, and to make sure that there is a common GND reference between the Analog Discovery 2 and the circuit under test, and that the common mode and differential voltages do not exceed specifications. Furthermore, | + | For safety reasons, it is the user’s responsibility to understand the powering and grounding scheme and to make sure that there is a common GND reference between the Analog Discovery 2 and the circuit under test, and that the common mode and differential voltages do not exceed specifications. Furthermore, |
<-- | <-- | ||
Line 77: | Line 77: | ||
The Analog Discovery 2 can be used with WaveForms' | The Analog Discovery 2 can be used with WaveForms' | ||
- | Each waveform generator channel is considered a single ended pin, however, a connected circuit must share a ground with the Analog Discovery 2. Each channel has a bandwidth of 8MHz through both the BNC and MTE connectors. AC amplitudes of +-5V and DC offsets of +-5V are supported. | + | Each waveform generator channel is considered a single ended pin, however, a connected circuit must share a ground with the Analog Discovery 2. |
Since the Analog Discovery 2's analog output channels are shared, the Waveform Generator instrument cannot be used at the same time as the Network Analyzer, or Impedance Analyzer instruments. | Since the Analog Discovery 2's analog output channels are shared, the Waveform Generator instrument cannot be used at the same time as the Network Analyzer, or Impedance Analyzer instruments. | ||
Line 137: | Line 137: | ||
{{test-and-measurement: | {{test-and-measurement: | ||
- | The Analog Discovery 2 can be used with WaveForms' | + | The Analog Discovery 2 can be used with WaveForms' |
Individual input/ | Individual input/ | ||
Line 179: | Line 179: | ||
The Analog Discovery 2 can be used with WaveForms' | The Analog Discovery 2 can be used with WaveForms' | ||
- | The Analog Discovery 2's digital input/ | + | The Analog Discovery 2's digital input/ |
Digital input/ | Digital input/ | ||
Line 186: | Line 186: | ||
==== Features ==== | ==== Features ==== | ||
- | * Virtual I/O devices (LEDs, buttons, switches & displays) | + | * Virtual I/O devices (LEDs, buttons, switches, & displays) |
* Customized visualization options available | * Customized visualization options available | ||
Line 201: | Line 201: | ||
The Analog Discovery 2 can be used with WaveForms' | The Analog Discovery 2 can be used with WaveForms' | ||
- | |||
- | Signals with minimum/ | ||
Since the Spectrum Analyzer instrument uses the same hardware resources as the Oscilloscope, | Since the Spectrum Analyzer instrument uses the same hardware resources as the Oscilloscope, | ||
Line 226: | Line 224: | ||
The Analog Discovery 2 can be used with WaveForms' | The Analog Discovery 2 can be used with WaveForms' | ||
- | Frequency sweeps can be performed in ranges between 1 mHz and 10 MHz with up to 10k samples per decade. The wave used for the sweep can be customized, and uses the same resources as the Waveform Generator instrument. | + | A sweep is performed in order to test the response of the circuit at varying frequencies. The wave used to perform this sweep can be customized and uses the same resources as the Waveform Generator instrument. |
The Network Analyzer instrument uses the analog output and analog input channels of the Analog Discovery 2 to probe a test circuit. The Network Analyzer can be configured to use an external signal to provide input to the circuit under test, rather than using the analog output channels. | The Network Analyzer instrument uses the analog output and analog input channels of the Analog Discovery 2 to probe a test circuit. The Network Analyzer can be configured to use an external signal to provide input to the circuit under test, rather than using the analog output channels. | ||
Line 246: | Line 244: | ||
The Analog Discovery 2 can be used with WaveForms' | The Analog Discovery 2 can be used with WaveForms' | ||
- | Frequency sweeps can be performed | + | A sweep is performed in order to test the response of the circuit at varying frequencies. The signal used to perform the sweep can be selected from a variety of presets, with configurable amplitude and offset. An external network analyzer reference circuit can be selected from a variety of options. |
The Impedance Analyzer instrument uses the analog output channels and analog input channels of the Analog Discovery 2 to probe a test circuit. | The Impedance Analyzer instrument uses the analog output channels and analog input channels of the Analog Discovery 2 to probe a test circuit. | ||
Line 283: | Line 281: | ||
{{test-and-measurement: | {{test-and-measurement: | ||
- | The Analog Discovery 2 can be used with WaveForms' | + | The Analog Discovery 2 can be used with WaveForms' |
Custom scripts can be written within the Protocol Analyzer instrument to generate sequences of SPI or I2C transactions. | Custom scripts can be written within the Protocol Analyzer instrument to generate sequences of SPI or I2C transactions. | ||
Line 304: | Line 302: | ||
Each of WaveForms' | Each of WaveForms' | ||
- | A variety of code examples are available in the application to aid in learning to write WaveForms scripts. Additional resources for writing scripts can be found on the Scopes | + | A variety of code examples are available in the application to aid in learning to write WaveForms scripts. Additional resources for writing scripts can be found on the Test and Measurement |
A plot pane within the Script instrument itself can be used to integrate data from many different instruments, | A plot pane within the Script instrument itself can be used to integrate data from many different instruments, | ||
Line 319: | Line 317: | ||
===== WaveForms Software Development Kit (SDK) ===== | ===== WaveForms Software Development Kit (SDK) ===== | ||
- | The WaveForms SDK is a set of software libraries and examples that can be used to develop custom applications that can control Digilent Test and Measurement devices. Supported languages include C, C++, C#, Visual Basic, and Python. Third party toolkits are available for LabVIEW and MATLAB. Instructions for using WaveForms with LabVIEW are available through the [[https:// | + | The WaveForms SDK is a set of software libraries and examples that can be used to develop custom applications that can control Digilent Test and Measurement devices. Supported languages include C, C++, C#, Visual Basic, and Python. Third party toolkits are available for LabVIEW and MATLAB. Instructions for using WaveForms with LabVIEW are available through the [[https:// |
==== Features ==== | ==== Features ==== |