====== OpenScope MZ - Calibration ====== | ====== OpenScope MZ - Calibration ====== | ||
+ | |||
+ | {{ youtube> | ||
The OpenScope MZ uses software calibration to improve measurement accuracy and each OpenScope MZ is calibrated as part of the manufacturing process. | The OpenScope MZ uses software calibration to improve measurement accuracy and each OpenScope MZ is calibrated as part of the manufacturing process. | ||
Line 10: | Line 12: | ||
- Launch [[http:// | - Launch [[http:// | ||
- | - [[/reference/software/waveforms-live/add-device | Add the OpenScope MZ]] to WaveForms Live. \\ | + | - [[software:waveforms-live:add-device| Add the OpenScope MZ]] to WaveForms Live. \\ |
- | - Click the **More** button on the device card and select **Configure**. \\ \\ {{ :reference:instrumentation: | + | - Click the **More** button on the device card and select **Configure**. \\ \\ {{ reference:test-and-measurement: |
- | - Under the **Calibration** section click **Calibrate**. \\ \\ {{ :reference:instrumentation: | + | - Under the **Calibration** section click **Calibrate**. \\ \\ {{ reference:test-and-measurement: |
- Device specific instructions are provided. | - Device specific instructions are provided. | ||
- Connect the solid red wire to the solid orange wire. | - Connect the solid red wire to the solid orange wire. | ||
- Connect the solid white wire to the solid blue wire. | - Connect the solid white wire to the solid blue wire. | ||
- | - Click **Next** to begin calibration. | + | - Click **Next** to begin calibration. |
- | - The calibration process takes about 30 seconds. | + | - The calibration process takes about 30 seconds. |
- | - The calibration information is displayed. | + | - The calibration information is displayed. |
- The new calibration data has been applied to the instrument and will be applied at boot. | - The new calibration data has been applied to the instrument and will be applied at boot. | ||