 ====== OpenScope MZ - Calibration ====== ====== OpenScope MZ - Calibration ======
 +
 +{{ youtube>Bxf24fgrgKM?large }}
  
 The OpenScope MZ uses software calibration to improve measurement accuracy and each OpenScope MZ is calibrated as part of the manufacturing process.  The OpenScope MZ **does not** need to be re-calibrated before each use, however, calibration is recommended before the first use and when changing environments, especially if there is a large temperature variation between environments.   The OpenScope MZ uses software calibration to improve measurement accuracy and each OpenScope MZ is calibrated as part of the manufacturing process.  The OpenScope MZ **does not** need to be re-calibrated before each use, however, calibration is recommended before the first use and when changing environments, especially if there is a large temperature variation between environments.  
Line 11: Line 13:
   - Launch [[http://waveformslive.com/ | WaveForms Live]].   - Launch [[http://waveformslive.com/ | WaveForms Live]].
   - [[/reference/software/waveforms-live/add-device | Add the OpenScope MZ]] to WaveForms Live. \\    - [[/reference/software/waveforms-live/add-device | Add the OpenScope MZ]] to WaveForms Live. \\ 
-  - Click the **More** button on the device card and select **Configure**. \\ \\ {{ :reference:instrumentation:openscope-mz:waveforms-live-openscope-calibration-4.png |}} \\ \\ {{ :reference:instrumentation:openscope-mz:waveforms-live-openscope-calibration-5.png |}} \\ \\ +  - Click the **More** button on the device card and select **Configure**. \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-4.png |}} \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-5.png |}} \\ \\ 
-  - Under the **Calibration** section click **Calibrate**. \\ \\ {{ :reference:instrumentation:openscope-mz:waveforms-live-openscope-calibration-6.png |}} \\ \\ +  - Under the **Calibration** section click **Calibrate**. \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-6.png |}} \\ \\ 
   - Device specific instructions are provided.    - Device specific instructions are provided. 
     - Connect the solid red wire to the solid orange wire.     - Connect the solid red wire to the solid orange wire.
     - Connect the solid white wire to the solid blue wire.     - Connect the solid white wire to the solid blue wire.
-    - Click **Next** to begin calibration.  \\ \\ {{ :reference:instrumentation:openscope-mz:waveforms-live-openscope-calibration-7.png |}} \\ \\  +    - Click **Next** to begin calibration.  \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-7.png |}} \\ \\  
-  - The calibration process takes about 30 seconds.  \\ \\ {{ :reference:instrumentation:openscope-mz:waveforms-live-openscope-calibration-8.png |}} \\ \\  +  - The calibration process takes about 30 seconds.  \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-8.png |}} \\ \\  
-  - The calibration information is displayed.  Select the desired **storage location** (flash is recommended) and click **Save**.  Then click **Done**.  \\ \\ {{ :reference:instrumentation:openscope-mz:waveforms-live-openscope-calibration-9.png |}} \\ \\ +  - The calibration information is displayed.  Select the desired **storage location** (flash is recommended) and click **Save**.  Then click **Done**.  \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-9.png |}} \\ \\ 
   - The new calibration data has been applied to the instrument and will be applied at boot.   - The new calibration data has been applied to the instrument and will be applied at boot.