Analog Discovery Pro 5000 Series Specifications

These specifications are valid following 30 minutes of warm-up and are typical at 25 °C unless otherwise noted.
Warranted specifications are valid at Tcal ±5 °C. Temperature coefficients are calculated using the temperature change from last external calibration.

Specifications not assigned to an individual device are valid for all Analog Discovery Pro 5000 series devices.

Mixed Signal Oscilloscope

Analog Input Channels

Supports the Oscilloscope, Voltmeter, Data Logger, Spectrum Analyzer, Network Analyzer, Impedance Analyzer, and Script Editor instruments.

Vertical System

ADP5250 ADP5470 ADP5490
Number of Channels 2 4 4
Input Type Single-ended, non-isolated
Connector Type Front panel BNC
Hardware Input Ranges
10 mV / div (100 mV peak to peak)
20 mV / div (200 mV peak to peak)
40 mV / div (400 mV peak to peak)
100 mV / div (1 V peak to peak)
200 mV / div (2 V peak to peak)
400 mV / div (4 V peak to peak)
1 V / div (10 V peak to peak)
2 V / div (20 V peak to peak)
4 V / div (40 V peak to peak)
4 mV / div (40 mV peak to peak)
10 mV / div (100 mV peak to peak)
20 mV / div (200 mV peak to peak)
40 mV / div (400 mV peak to peak)
100 mV / div (1 V peak to peak)
200 mV / div (2 V peak to peak)
400 mV / div (4 V peak to peak)
1 V / div (10 V peak to peak)
2 V / div (20 V peak to peak)
4 V / div (40 V peak to peak)
Resolution 8 bits
Absolute Resolution up to 0.391 mV (at 10 mV/div range) up to 0.156 mV (at 4 mV/div range)
Accuracy (warranted) ±2% of input ± 1% full scale (V peak to peak)
Bandwidth (-3dB)1) 100 MHz 350 MHz 500 MHz
Hardware Bandwidth Limit (-3dB) 20 MHz, can be enabled or disabled in software per channel
Input Impedance 1 MΩ ││ 20 pF 1 MΩ ││ 15 pF or 50 Ω2) 1 MΩ ││ 12 pF or 50 Ω3)
Input Coupling AC, DC
Vertical Sensitivity (range) 200 µV/div to 4 V/div (10 divisions)4)
Acquisition Modes Average, decimate, min/max

DC Offset Range

Hardware Range Values Full Scale Programmable Offset Range
40 mV/div and smaller 10 V peak-to-peak ±5 V
100 mV/div and larger 40 V peak-to-peak ±20 V

1) A probe with the appropriate frequency response is required to reach the specified oscilloscope bandwidth.
2), 3) Maximum voltage when using 50 Ω input mode is 5 V RMS. For a periodic waveform with frequency below 100 kHz, the maximum voltage is derated to 2.5 V RMS.
4) Divisions in this context are the ten horizontal strips in the Analog Input graph windows within WaveForms. Vertical sensitivity specifies the height of one strip in the plot.

Horizontal System

ADP5250 ADP5470 ADP5490
Maximum Sample Rate 1 GS/s single channel
500 MS/s per channel, dual channel
1.5 GS/s per channel 2 GS/s per channel
Maximum Record Length 1 MS per channel

Note: Acquisitions are made simultaneously on all enabled channels in all modes.


Digital Input Channels

Supports the Logic Analyzer and Script Editor instruments.

Vertical System

Number of Channels 34
Connector 100 mil 2×20 MTE
Input Voltage 0 V to 5 V1)
Input Current ≤50 µA
Input Threshold Adjustable, 0 V to 2 V
Threshold Accuracy 350 mV
Input Impedance 100 kΩ ││ 7.5 pF (nominal) pulled to -2.0 V to +6.5 V, varies with the input threshold setting
Logic Analyzer Interpreters SPI, I2C, UART, CAN, I2S, 1-Wire, PS/2, HDMI CEC, Manchester codes, JTAG, GPIB, SWD, custom, and more

1) Mixed signal oscilloscope digital channels are designed to withstand accidental overvoltage from signals on all three of the Analog Discovery Pro 5000 series devices. They are not recommended for use with signals likely to exceed 0 V to 5 V in normal operation.

Horizontal System

Maximum Sample Rate (warranted) 1 GS/s (down to ~15 kS/s)
Record Length (Typical) 1 MS
Record Length (Minimum)1) 4 kS

1) Under most conditions, the logic analyzer can acquire 1 MS of data. Under some conditions with very high sustained activity on multiple inputs, the logic analyzer may only capture 4 kS of data.

Arbitrary Waveform Generator (Wavegen)

Supports the Waveform Generator, Network Analyzer, Impedance Analyzer, and Script Editor instruments.

Vertical System

ADP5250 ADP5470 ADP5490
Number of Channels 1
Output Type Single-ended
Connector Type Front panel BNC
Standard Functions Sine, square, triangle, ramp up, ramp down, DC voltage, noise, pulse, others
Output Voltage Range1) ±6 V at 50 Ω output impedance
±12 V at high (>10 kΩ) output impedance2)
Resolution 14 bits
Absolute Resolution 0.732 mV at 50 Ω output impedance
1.46 mV at high (>10 kΩ) output impedance3)
Accuracy ±(1% of output value ±5 mV)4)
Output Impedance 50 Ω
Bandwidth ±0.3 dB to 20 MHz ±0.3 dB to 40 MHz
Switchable filter 36 MHz lowpass, 7 pole, elliptical filter 71 MHz lowpass, 7 pole, elliptical filter
Total Harmonic Distortion (1 MHz) -55 dBC
Total Harmonic Distortion (10 MHz) -50 dBC
Phase Noise -125 dBc/Hz at 10 kHz offset
Rise/fall time <20 ns (10% to 90%)
Overshoot <5%5)
Jitter 8 ns cycle-to-cycle
Overvoltage Protection Short-circuit protected
Custom Waveform Files Supported Import files *.csv, *.txt, *.mp3, *.wav, *.wmv & *.avi, export as image, or as raw data in *.csv, *.txt or *.tdms formats

DC Output Range

Output Impedance Offset Range6)
50 Ω ±6 V
High (>10 kΩ) output impedance7) ±12 V

1), 6) The combination of signal amplitude and DC offset cannot exceed the output range specifications.
2), 3), 7) The high output impedance is a load applied by the user to the AWG output
4) Into a >10 kΩ load applied by the user
5) At 5 MHz and slower Square wave

Horizontal System

ADP5250 ADP5470 ADP5490
Maximum Sample Rate 125 MS/s 200 MS/s
Buffer Size Up to 1 MS

Digital I/O

Digital I/O channels

Number of Channels 8
Connector 1×14 Terminal Block
Function Control Individually programmable as Digital I/O
Input Voltage 0 V to 5 V TTL1)
Output Voltage 3.3 V LVTTL
Drive Strength 4 mA
Pull Resistors 10 kΩ pull-down to 0 V for all I/O pins
1.5 kΩ pull-up to 3.3 V on I/O 6 and 7 when I2C is used

External Power

Number of Channels 1
Connector 1×14 Terminal Block
Voltage 3.3 V ± 10%
Current 20 mA

1) Digital I/O lines are designed to withstand accidental overvoltage from digital signals on Digilent's Test and Measurement offerings or similar devices. They are not recommended for use with signals likely to exceed 0 V to 5 V in normal operation.

Trigger System

Trigger Features

ADP5250 ADP5470 ADP5490
System Trigger Sources Oscilloscope analog channels, Oscilloscope digital channels, Function generator start, External trigger (TRIG1), power line frequency
Trigger Modes None, Auto, Normal, Manual (Forced Trigger), Single
Analog Trigger Edge, level, hysteresis
Digital Trigger Edge, level, pattern, glitch
Oscilloscope Trigger Resolution 1 ns 667 ps 500 ps
Logic Analyzer Trigger Resolution 1 ns 1 ns 1 ns
Wavegen Output Trigger Resolution 8 ns 5 ns
Output Trigger Sources Oscilloscope Start, Waveform Generator Start, Power line frequency, Drive low, Drive high

External Trigger Characteristics

Number of Channels 1
Trigger Type Digital
External Trigger Connector Front panel BNC
Input Voltage 5 V compatible TTL1)
Output Type 3.3 V LVTTL output
Channel Pull Resistors 1 MΩ pull-down
Direction Control Individually programmable as input or output

1) The external trigger line is designed to withstand accidental overvoltage from digital signals on Digilent's Test and Measurement offerings or similar devices. They are not recommended for use with signals likely to exceed 0 V to 5 V in normal operation.

Digital Multimeter

Functionality

Functions DC Voltage, AC Voltage, DC current, AC current, resistance, diode, continuity
Resolution 5½ digits
Sample Rate 5 S/s

Caution: Do not use this device for connection to signals or for measurements within Measurement Categories III or IV. For more information about Measurement Categories, refer to the Safety Voltages section.

Input Protection

ADP5250 ADP5470 ADP5490
Resistance, Diode Up to 300 V DC
DC and AC Voltage Up to 300 V DC or 265 V AC RMS, 400 V AC (peak)
DMM A Current Connector Fuse1)2) Internal ceramic fuse, 10 A 250 V, time-delay, 5 x 20 mm, T 10A H 250V (Bussman part number D505H-10-R at www.cooperindustries.com) Internal ceramic fuse, 11 A, 1 kV AC, 10.3 × 38 mm, F 11A 1000V (SIBA part number 5019906.11 at www.siba-fuses.com)
DMM mA Current Connector Fuse3)4) Internal ceramic fuse, 1.25 A 250 V, time-delay, 5 x 20 mm, T 1.25A H 250V (Bussman part number S505H-1.25-R at www.cooperindustries.com) Internal ceramic fuse, 1 A, 500 V AC, 5 × 20 mm, T 1A H 400V (Littelfuse part number 0477001.MXP at www.littelfuse.com)
Maximum common-mode voltage 300 V DC or AC (rms)

1), 3) Fuses are located on bottom of device underneath door. Use Phillips #1 screwdriver for removal. Ensure all hazardous voltages are disconnected from the device prior to removal of door.
2), 4) When the fuse symbol, , is marked on a device, take proper precautions.

DC

DC Voltage Accuracy

Range Input Impedance 1-Year Accuracy ±(% of Reading + % of Range) (warranted) Temperature Coefficient ±(% of Reading + % of Range)/C° (warranted)
100 mV1) <10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005
1 V >10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005
10 V >10 GΩ, 10 MΩ 0.015 +0.005 0.001 + 0.0005
100 V 10 MΩ 0.035 + 0.005 0.005 + 0.0005
300 V 10 MΩ 0.035 + 0.005 0.005 + 0.0005

1) Add 15 µV if not immediately following offset null.

DC Current Accuracy

Range Burden Voltage 1-Year Accuracy ±(% of Reading + % of Range) (warranted) Temperature Coefficient ±(% of Reading + % of Range)/C° (warranted)
10 mA <0.03 V 0.070 + 0.020 0.0035 + 0.0010
100 mA <0.3 V 0.070 + 0.003 0.0020 + 0.0010
1 A <0.03 V 0.130 + 0.025 0.0065 + 0.0010
10 A1) <0.3 V 0.130 + 0.004 0.0045 + 0.0010

1) 30 seconds on, 30 seconds off. Add 300 ppm for currents >2.2 A. After measuring >5 A, wait two minutes to get full accuracy in the 1 A range.

DC Resistance Accuracy (2-Wire), 1 V Open Circuit Voltage

Range Short-Circuit Current 1-Year Accuracy ±(% of Reading + % of Range)1) (warranted) Temperature Coefficient ±(% of Reading + % of Range)/C° (warranted)
100 Ω 170 µA 0.018 + 0.050 0.0010 + 0.0005
1 kΩ 170 µA 0.018 + 0.005 0.0010 + 0.0005
10 kΩ 70 µA 0.018 + 0.005 0.0010 + 0.0005
100 kΩ 10 µA 0.018 + 0.005 0.0010 + 0.0005
1 MΩ 1.1 µA 0.035 + 0.005 0.0040 + 0.0005
10 MΩ 1.1 µA 0.150 + 0.005 0.0100 + 0.0005
100 MΩ 1.1 µA 1.3 + 0.005 0.1000 + 0.0005

1) Perform offset nulling for these values

Caution: The input terminals of the DMM are not protected for electromagnetic interference. As a result, the DMM may experience reduced measurement accuracy or other temporary performance degradation when connected to unshielded test leads in an environment with radiated or conducted radio frequency electromagnetic interference.

Additional DC Characteristics

DC Continuity Accuracy Range 100 Ω
DC Diode Test Range 2 V
Effective Common-Mode Rejection-Mode Ratio (CMRR), 1 kΩ resistance in LO lead >100 dB
Normal-Mode Rejection Ratio (NMRR), 50/60 Hz ± 0.1% >100 dB
Overrange 105% of range except 300 V

AC

AC Voltage Accuracy

Range (rms) Peak Voltage Frequency 1-Year Accuracy ±(% of Reading + % of Range) (warranted) Temperature Coefficient ±(% of Reading + % of Range)/C° (warranted)
100 mV, 1 V, 10 V,
100 V, 265 V
±210 mV, ±2.1 V, ±21 V,
±210 V, ±400 V
20 Hz to 45 Hz 0.91 + 0.10 0.01 + 0.005
45 Hz to 65 Hz 0.30 + 0.05 0.01 + 0.005
65 Hz to 1 kHz 0.21 + 0.05 0.01 + 0.005
1 kHz to 5 kHz 0.12 + 0.05 0.01 + 0.005
5 kHz to 20 kHz 0.35 + 0.05 0.01 + 0.005

AC Current Accuracy

Range (rms) Short-Circuit Current Burden Voltage (rms) Frequency 1-Year Accuracy ±(% of Reading + % of Range) (warranted) Temperature Coefficient ±(% of Reading + % of Range)/C° (warranted)
5 mA ±10.5 mA <0.02 20 Hz to 1 kHz 0.20 + 0.01 0.01 + 0.005
5 mA ±10.5 mA <0.02 1 kHz to 5 kHz 0.60 + 0.01 0.01 + 0.005
50 mA ±105 mA <0.2 V 20 Hz to 1 kHz 0.20 + 0.01 0.01 + 0.005
50 mA ±105 mA <0.2 V 1 kHz to 5 kHz 0.50 + 0.01 0.01 + 0.005
500 mA ±1.05 A <0.02 V 20 Hz to 1 kHz 0.15 + 0.01 0.01 + 0.005
500 mA ±1.05 A <0.02 V 1 kHz to 5 kHz 0.50 + 0.01 0.01 + 0.005
5 A ±1.05 A <0.02 V 20 Hz to 1 kHz 0.25 + 0.03 0.01 + 0.005
5 A ±1.05 A <0.02 V 1 kHz to 5 kHz 0.60 + 0.03 0.01 + 0.005

Caution: The input terminals of the DMM are not protected for electromagnetic interference. As a result, the DMM may experience reduced measurement accuracy or other temporary performance degradation when connected to unshielded test leads in an environment with radiated or conducted radio frequency electromagnetic interference.


Additional AC Characteristics

Input Impedance 10 MΩ ││ 200 pF
Effective Common-Mode Rejection-Mode Ratio (CMRR), 1 kΩ resistance in LO lead >70 dB (DC to Hz)

Programmable Power Supply

ADP5250 Characteristics

+6 V +25 V -25 V
DC Voltage Output (warranted) 0 V to +6 V 0 V to +25 V 0 V to -25 V
DC Current Output (warranted)1)2) 1 A 500 mA (isolated) 500 mA (isolated)
Voltage Programming Accuracy3), ±(% of reading + offset) (warranted) 0.1% + 5 mV 0.15% + 20 mV 0.15% + 20 mV
Current Programming Accuracy4)5), ±(% of reading + offset) (warranted) 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Voltage Readback Accuracy6), ±(% of reading + offset) (warranted) 0.1% + 5 mV 0.15% + 20 mV 0.15% + 20 mV
Current Readback Accuracy7), ±(% of reading + offset) (warranted) 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Voltage Programming Resolution 1.7 mV 6.5 mV 6.5 mV
Current Programming Resolution 0.30 mA 0.15 mA 0.15 mA
Voltage Readback Resolution 0.41 mV 1.7 mV 1.7 mV
Current Readback Resolution 70 µA 35 µA 35 µA
Load Regulation8) ±(% of reading + offset) 0.01% + 25 mV 0.03% + 5 mV 0.03% + 5 mV
Overvoltage Protection 10 V 30 V
Reverse Voltage Protection Reverse clamp diode, protected by self-resetting fuse
Connector Type 1×6 Pluggable Screw Terminal Block

ADP5470 and ADP5490 Characteristics

+6 V +25 V -25 V
DC Voltage Output (warranted) 0 V to +6 V 0 V to +25 V 0 V to -25 V
DC Current Output (warranted)9)10) 3 A 1 A (isolated) 1 A (isolated)
Voltage Programming Accuracy11), ±(% of reading + offset) (warranted) 0.1% + 5 mV 0.1% + 20 mV 0.1% + 20 mV
Current Programming Accuracy12)13), ±(% of reading + offset) (warranted) 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Voltage Readback Accuracy14), ±(% of reading + offset) (warranted) 0.1% + 5 mV 0.1% + 20 mV 0.1% + 20 mV
Current Readback Accuracy15), ±(% of reading + offset) (warranted) 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Voltage Programming Resolution 1.6 mV 6.6 mV 6.6 mV
Current Programming Resolution 0.90 mA 0.30 mA 0.30 mA
Voltage Readback Resolution 0.40 mV 1.7 mV 1.7 mV
Current Readback Resolution 210 µA 70 µA 70 µA
Load Regulation16) ±(% of reading + offset) 0.01% + 25 mV 0.03% + 5 mV 0.03% + 5 mV
Overvoltage Protection 30 V
Reverse Voltage Protection Reverse clamp diode, protected by self-resetting fuse
Connector Type 6 banana connectors

1), 4), 9), 12) Minimum programmable current limit is 1% of range
2), 10) The +25 V and –25 V channels are bank isolated from ground but not from each other.
3), 5), 6), 7), 11), 13), 14), 15) Programming and readback accuracy specified at no load.
8), 16) Change in output voltage for any load within range.

Features

Spectrum Analyzer

ADP5250 ADP5470 ADP5490
Frequency Range1) 0 Hz to 100 MHz, up to 10,001 samples 0 Hz to 350 MHz, up to 10,001 samples 0 Hz to 500 MHz, up to 10,001 samples
Display Modes Magnitude, average, peak hold, min hold, count, and more
Y Axis Peak (V), Ṽ/√Hz, dBṼ, dBu, dBm, dBFS, and more
X Axis Linear or Logarithmic
Windowing Functions Rectangular, Triangular, Hamming, Hann, Cosine, Blackman-Harris, Flat Top, Kaiser, and more

1) Higher frequencies may be selected within WaveForms, but results may be limited by the analog input bandwidth of the hardware.

Network Analyzer

ADP5250 ADP5470 ADP5490
Frequency Range1) 20 Hz to 100 MHz, up to 10,001 samples 10 Hz to 350 MHz, up to 10,001 samples 10 Hz to 500 MHz, up to 10,001 samples
Display Modes Magnitude, Phase
Y Axis Linear or Logarithmic, dB, gain, Vpeak, and more magnitude representations
X Axis Linear or Logarithmic
Plots Bode, Time, FFT, Nichols, Nyquist

1) Higher frequencies may be selected within WaveForms, but results may be limited by the analog input and output bandwidth of the hardware.

Impedance Analyzer

ADP5250 ADP5470 ADP5490
Frequency Range 20 Hz to 20 MHz, up to 10,001 steps 10 Hz to 20 MHz, up to 10,001 steps 10 Hz to 40 MHz, up to 10,001 steps
Measurements Magnitude, Phase, Impedance, Admittance, Inductance, Capacitance, and more
Y Axis Linear or Logarithmic
X Axis Linear or Logarithmic
Plots Bode, Time, FFT, Nichols, Nyquist, Custom
Measurements Impedance, Admittance, Inductance, Capacitance, and more

Protocol Analyzer

Shares digital input/output channels with Digital I/O: See the Digital I/O specifications.

Protocols1) SPI and I2C, host controller mode and custom only

1) This functionality is implemented by WaveForms software on the host system.

Math Channels

Operations Addition “+”, Subtraction “-”, Multiplication “*”, Division “/”, Remainder “%”
Brackets Parenthesis “()”, Square “[]”
Constants Exp, Ln, Log, Pi
Functions Logarithm, power, minimum, maximum, square root, sine, cos, tan, arccos, arctan, arctan2, absolute value, round, floor, ceil, random, and more
Operands All analog and digital input channels, reference waveforms, time, constants, Pi

Device Configurations

Each Analog Discovery Pro 5000 Series device has a single device configuration that allocates predefined buffer sizes for the analog input, analog output, and digital logic systems.

Table 1: Analog Discovery Pro (ADP5000) Series Configuration Option
Configuration Scope Buffer Size Wavegen Buffer Size Logic Buffer Size
1 One million points per channel 64 kS for each channel, one million point buffer when used in play mode One million samples aggregate

Note: Memory sizes, including buffer sizes, specified in units like kS and MS, are rounded from equivalent binary power units, such as MiS. For example, a listed 64 MS is rounded from 64 MiS, which is 67,108,864 samples.

Connectivity

USB Interface

Device to Computer Interface USB 2.0 Hi-Speed, Type B

Ethernet Interface

ADP5250 ADP5470 ADP5490
Network Interface N/A 1000 Base-TX, full-duplex
100 Base-TX, full-duplex
100 Base-TX, half-duplex
10 Base-T, full-duplex
10 Base-T, half-duplex
Communication Rates N/A 10/100/1000 Mbps, auto-negotiated
Connector Type N/A RJ-45

Note: The MAC address of the Ethernet port is listed on the sticker located on the underside of the device.


Power Requirements

The ADP5000 series devices require external power.

ADP5250 ADP5470 ADP5490
Voltage Input Range 100 VAC to 240 VAC, 50/60 Hz
Power Consumption 100 W maximum 150 W maximum
Power Input Connector IEC C13 power connector
Power Disconnect The AC power cable provides main power disconnect. Do not position the equipment so that it is difficult to disconnect the power cable. Depressing the front panel power button does not inhibit the internal power supply.

Physical Characteristics

Dimensions

ADP5250 ADP5470 ADP5490
Dimensions (Width x Depth x Height) 25.40 cm x 19.05 cm x 7.77 cm
(10.00 in x 7.50 in x 3.06 in)
30.48 cm × 20.32 cm × 9.40 cm
(12.0 in. × 8.0 in. × 3.7 in.)
Weight 2.05 kg (4 lb 8.3 oz) 3.130 kg (6 lb 4.4 oz)

Note: If the device needs to be cleaned, wipe with a dry towel.

Note: Use the ADP5000 devices in a horizontal orientation. Allow at least 10.16 cm (4.0 in.) of clearance in front and behind the device for USB, power, and common connector cabling.

Connectivity

ADP5250 ADP5470 ADP5490
Mixed Signal Oscilloscope Two BNC Four BNC
Logic Analyzer One 2×20 shrouded IDC header
External Trigger One BNC
Arbitrary Waveform Generator One BNC
Digital I/O Type One Pluggable screw terminal, 3.5 mm (14 position)
Digital I/O Screw Terminal Wiring 0.1 mm2 to 2.0 mm2 (30 AWG to 14 AWG)
Digital IO Screw Terminal Torque 0.25 N · m (2.2 lb · in.)
Digital Multimeter Four 4 mm banana jacks
DC Power Supply Type One Pluggable screw terminal, 3.81 mm (6 position) Six 4 mm binding posts
DC Power Supply Screw Terminal Wiring 0.1 mm2 to 2.0 mm2 (30 AWG to 14 AWG) N/A
DC Power Supply Screw Terminal Torque 0.25 N · m (2.2 lb · in.) N/A
Security Cable Slot 1, complies with Kensington security slot dimensions

Safety Voltages

Connect only voltages that are within these limits.

DMM Isolation Voltages (Ground to Earth)

Hazardous Voltage: This icon denotes a warning advising you to take precautions to avoid electrical shock.

Channel-to-earth Ground

Continuous 300 V, Measurement Category II1)
Withstand 3000 V (RMS), verified by a 5 s dielectric withstand test

1) Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

Caution: Do not connect the ADP5000 series hardware to signals or use for measurements within Measurement Categories III or IV.

DC Power Supply Isolation Voltages

+25V and -25V to earth ground, continuous 60 VDC, Measurement Category 11)

1) Measurement Categories CAT I and CAT O (Other) are equivalent. The input circuits are not intended for direct connection to the MAINs building installations of Categories CAT II, CAT III, or CAT IV.

Environmental

Operating Temperature 0 °C to 40 °C
Storage Temperature -20 °C to 70 °C
Operating Humidity 10% to 90% RH, non-condensing
DMM full accuracy at 10% to 80%
Storage Humidity 5% to 95% RH, non-condensing
Cooling Forced air circulation (positive pressurization) through a fan. Fan speed automatically adjusts according to operating conditions. Intake and exhaust locations are on rear of device. Ensure that the intake and exhaust locations are not obstructed.
Pollution Degree 2
Maximum Altitude 2,000 m

Note: For indoor use only.


Shock and Vibration

Operational Shock 30 g peak, half sine, 11 ns pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)
Random Vibration (Operating) 5 Hz to 500 Hz, 0.3 grms
Random Vibration (Nonoperating) 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

Certifications