Starting with the Analog Discovery

1. Hardware

To set up the Analog Discovery:

  1. First, install the Digilent WaveForms 3 software on your PC. See Installer Details for more information.
  2. Connect the Analog Discovery to your PC using a USB cable.

Pin-out:

Recommended Operating Conditions Absolute Maximum Ratings
Oscilloscope (1+,1-,2+,2-) ±50 V differential, ±25 V single ended ±50 V
WaveGen (W1,W2) ±5 V, 10 mA ±5.8 V, 50 mA
Power Supplies (V+,V-) +5 V, -5 V, 100 mA +5.2 V, -5.2 V, 200 mA
Triggers, Digital IO (T1,T2,0,1,2,…) LVCMOS 3.3 V, 4 mA ± 20 V

2. WaveForms Software

See Installer Details for information on installing WaveForms.

See WaveForms for information on using the software.

3. Troubleshooting

In case you receive the error message “Communication with device failed” or “Device configuration failed”:

The Analog Discovery's system monitor displays the voltage, current, and temperature. In the main window, click the button in the status bar to show this information.

When no instrument is running, the device power consumption is around 2 watts (400mA). Depending on the usage, consumption can increase subject to the following limitations:

If you receive the previously mentioned error message(s), or if the voltage is less than 4.5 V, try the following options:

4. Oscilloscope

The Analog Discovery has two differential oscilloscope input channels.

Specifications:

Low gain High gain
Range > 500 mV/div ⇐ 500 mV/div
Differential input voltage N/A -5 V to +5 V
Common mode input voltage N/A -20 V to +5 V
Positive and negative input voltage to GND ±25 V N/A

5. Arbitrary Waveform Generator

The Analog Discovery is equipped with two Arbitrary Waveform Generators channels. Verify the voltage on the Waveform Generator channels before connecting it to a circuit.

Specifications:

6. Power Supplies

The Analog Discovery has two fixed power supplies. NOTE: When the power supplies need to drive high capacitive loads, start the supply before connecting the load.

Specifications:

7. Digital I/O

The Analog Discovery has 16 digital I/O (DIO 0-15) and 2 trigger I/O pins.

Specifications:

8. Power Requirements and Limitations

Power to the Analog Discovery is provided by the USB port of the computer to which it is connected. In order to protect the computer's USB port, the current drawn by the Analog Discovery is limited to less than 600 mA. Attempts to draw more power than this from the USB port will cause the Analog Discovery to shut down and display an error message.

The voltage available from the USB port and the current being consumed by the device are displayed on the System Monitor in the lower-right corner of the WaveForms main window; an example is shown below.

The Analog Discovery provides four voltage supplies which can be used to provide power to circuits being tested. These are: V+ and V- in the Voltage instrument, and channels 1 and 2 of the waveform generator.

Under normal operation, with no instruments in use and no load applied to the voltage supplies, the Analog Discovery draws about 400 mA of current from the USB port.

The allocation of current to the various voltage supplies is left up to the user as much as possible. This allows the user to tailor the available power to their desired uses, which increases the utility of the device. There are, however, some limitations placed on the current provided: