====== Analog Discovery Pro (ADP5250) Specifications ====== /* Additional material for ADP5470 and ADP5490. https://www.ni.com/docs/en-US/bundle/vb-8034-specs/page/specs.html https://www.ni.com/en-us/shop/model/vb-8034.html https://www.ni.com/docs/en-US/bundle/vb-8054-specs/page/specs.html https://www.ni.com/en-us/shop/model/vb-8054.html PDFs of NI publicly published Specifications, GSGs, Calibration procedures, Visual Diagrams, and various Regulatory material have all been downloaded to James' computer Dec 18th 2023. */ ~~TechArticle~~ These specifications are valid following 30 minutes of warmup and are typical at 25 °C unless otherwise noted. ===== Mixed Signal Oscilloscope ===== ==== Analog Input Channels ==== Supports the Oscilloscope, Data Logger, Spectrum Analyzer, Network Analyzer, Impedance Analyzer, and Script Editor instruments. === Vertical System === |<100% 25% >| ^ Number of Channels | 2 single ended, non-isolated || ^ Bandwidth (-3 dB) | 100 MHz || ^ Resolution | 8 bits || ^ Accuracy | ±2% of input ±1% full scale (V peak to peak) || ^ Input Coupling | AC, DC || ^ Vertical Sensitivity (range) | 10 mV / div (100 mV peak to peak) \\ 20 mV / div (200 mV peak to peak) \\ 50 mV / div (400 mV peak to peak) \\ 100 mV / div (1 V peak to peak) \\ 200 mV / div (2 V peak to peak) \\ 500 mV / div (4 V peak to peak) \\ 1 V / div (10 V peak to peak) \\ 2 V / div (20 V peak to peak) \\ 5 V / div (40 V peak to peak)|| ^ Input Impedance | 1 MΩ ││ 20 pF || ^ Acquisition Modes | Sample, peak detect, averaging || ^ DC Offset Range | 10 mV/div, 20 mV/div, 50 mV/div ranges: ±5 V \\ 100 mV/div, 200 mV/div, 50 mV/div, 1 V/div, 2 V/div, 5 V/div ranges: ±20 V || ---- === Horizontal System === |<100% 25%>| ^ Maximum Sampling Rate | 1 GS/s single channel \\ 500 MS/s per channel, dual channel || ^ Maximum Record Length | 1 MS per channel || ---- ==== Digital Channels / Logic Analyzer ==== Supports the Logic Analyzer and Script Editor instruments. {{ :test-and-measurement:analog-discovery-pro-5250:connector.png?400 |}} === Vertical System === |<100% 40%>| ^ Number of Channels | 34 [(At time of writing, WaveForms software does not support use of the CLK0/CLK1 lines. All 32 DIO pins are fully supported.)] || ^ Maximum Input Frequency | 100 MHz || ^ Input Voltage | 0 V to 5 V || ^ Input Current | ≤50 µA || ^ Input Threshold | Programmable, 0 V to 2.0 V || ^ Threshold Accuracy | 350 mV || ^ Input Impedance | 100 kΩ ││ 7.5 pF pulled to -2.0 V to +6.5 V, varies with input threshold setting || ^ Additional / Internal Channels | Digital I/O lines, function generator start, external trigger (TRIG), power line frequency || **Note:** Mixed signal oscilloscope digital channels are designed to withstand accidental overvoltage from signals on the ADP5250. They are not recommended for use with signals likely to exceed 0 V to 5 V in normal operation. ~~REFNOTES~~ === Horizontal System === |<100% 40%>| ^ Timing Mode Sample Rate | 1 GS/s (down to ~15 kS/s) || ^ Maximum External Sample Clock Rate | 100 MHz || ^ Record Length (Typical) | 1 MS || ^ Record Length (Minimum) | 4 kS || ^ Decimation | External Sample Clock, 1:1, 2:1, and n*4:1 where n is an integer || ^ Maximum Sample Compression | 215 to 1 || ---- ==== Triggering ==== |<100% 20% 20%>| ^ Trigger Modes || Normal, Auto, Force, Single | ^ Trigger Sources || Oscilloscope analog channels \\ Oscilloscope digital channels \\ Function generator start \\ Digital I/O lines \\ External trigger (TRIG) \\ Power line frequency | ^ Trigger Types ^ Analog | Edge with hysteresis | ^ ::: ^ Digital | Edge, pattern, glitch, level | ^ Trigger Resolution ^ Analog / Oscilloscope | 667 ps | ^ ::: ^ Digital / Logic Analyzer | 1 ns | ^ Trigger Export || Available through external trigger (TRIG) | ---- ===== Waveform Generator (WaveGen) ===== Supports the Waveform Generator, Network Analyzer, Impedance Analyzer, and Script Editor instruments. |<100% 25% 25%>| ^ Waveforms || Sine, square, ramp / triangle, DC, arbitrary | ^ Update Rate || 125 MS/s | ^ Resolution || 14 bits | ^ Number of Channels || 1 | ^ Output Impedance || 50 Ω | ^ Switchable Filter || 36 MHz lowpass, 7-pole, elliptical | ^ Sine ^ Maximum Frequency | 20 MHz | | ::: ^ Total Harmonic Distortion (1 MHz) | -55 dBc | | ::: ^ Total Harmonic Distortion (10 MHz) | -50 dBc | | ::: ^ Spurious Free Dynamic Range (SFDR) | -70 dB at 1 MHz (non-harmonic) | | ::: ^ Phase Noise (1 MHz) | -115 dBc/Hz at 10 kHz offset | ^ Square ^ Maximum Frequency | 20 MHz | | ::: ^ Rise / Fall Time | <20 ns (10% to 90%) | | ::: ^ Overshoot | <5% | | ::: ^ Jitter | 8 ns cycle-to-cycle | ^ Ramp / Triangle Maximum Frequency || 20 MHz | ^ Accuracy (with >10 kΩ load) ^ Amplitude (1 kHz sine) | ±(1% of output value ±5 mV) | | ::: ^ DC | ±(1% of output value ±7.5 mV) | ^ Output Range ^ 50 Ω | ±6 V | | ::: ^ Hi-Z (>10 kΩ) | ±12 V | ^ DC Offset ^ 50 Ω | ±6 V | | ::: ^ Hi-Z (>10 kΩ) | ±12 V | ^ Frequency ^ Accuracy | ≤100 ppm | | ::: ^ Resolution | 1 µHz | ^ Arbitrary Waveforms ^ Points | 1 MS | | ::: ^ Sample Rate | 125 MS/s | ^ Protection || Short-circuit protected | **Note:** The combination of signal amplitude and DC offset cannot exceed the output range specifications. The impedances listed are the loads applied by the user to the Waveform Generator output. ---- {{ :test-and-measurement:analog-discovery-pro-5250:connector2.png?400 |}} ===== Digital I/O ===== Supports the Static I/O and Script Editor instruments. |<100% 40%>| ^ Number of Channels | 8 || ^ Direction Control | Input or output, software-selectable || ^ Logic Level | 5 V compatible TTL input \\ 3.3 V LVTTL output || ^ Drive Strength | 4 mA || ^ Input Voltage | 0 to 5 V || ^ DIO Channel Pull Resistors | 10 kΩ, pull-down on dig/<0..7> \\ 10 kΩ, configurable pull-up to 3.3 V on dig/<6,7> || **Note:** Digital I/O lines are designed to withstand accidental overvoltage from signals on the ADP5250 or similar devices. They are not recommended for use with signals likely to exceed 0 V to 5 V in normal operation. ---- ====Digital Power Supply==== |<100% 40%>| ^ Voltage | 3.3 V ± 10% | ^ Current | 20 mA | ---- ===== Digital Multimeter ===== |<100% 40%>| ^ Functions | DC Voltage, AC Voltage, DC current, AC current, resistance, diode, continuity || ^ Resolution | 5½ digits || ^ Sample Rate | 5 S/s || **Caution:** Do not use this device for connection to signals or for measurements within Measurement Categories III or IV. For more information about Measurement Categories, refer to the [[#safety_voltages|Safety Voltages]] section. === Input Protection === |<100% 20% 25%>| ^ Resistance, Diode || Up to 300 V DC | ^ DC and AC Voltage || Up to 300 V DC or 265 V AC (rms) 400 V AC (peak) | ^ DC and AC Current ^ DMM A Current Connector Fuse | Internal ceramic fuse, 10 A 250 V, time-delay, 5 x 20 mm, T 10A H 250V (Bussman part number D505H-10-R at www.cooperindustries.com) | ^ ::: ^ DMM mA Current Connector Fuse | Internal ceramic fuse, 1.25 A 250 V, time-delay, 5 x 20 mm, T 1.25A H 250V (Bussman part number S505H-1.25-R at www.cooperindustries.com) | **Caution:** Fuses are located on bottom of device underneath door. Use Phillips #1 screwdriver for removal. Ensure all hazardous voltages are disconnected from the device prior to removal of door. **Fuse:** When the fuse symbol is marked on a device, take proper precautions. |<100% 40%>| ^ Maximum common-mode voltage | 300 V DC or AC (rms) || ---- ====DC==== === DC Voltage Accuracy === |<100% 25%>| ^ **Range** | **Input Impedance** | **1-Year Accuracy ±(% of Reading + % of Range)** | ** Temperature Coefficient ±(% of Reading + % of Range)/C°** || ^ 100 mV* | <10 GΩ, 10 MΩ | 0.015 + 0.005 | 0.001 + 0.0005 || ^ 1 V | >10 GΩ, 10 MΩ | 0.015 + 0.005 | 0.001 + 0.0005 || ^ 10 V | >10 GΩ, 10 MΩ | 0.015 +0.005 | 0.001 + 0.0005 || ^ 100 V | 10 MΩ | 0.035 + 0.005 | 0.001 + 0.0005 || ^ 300 V | 10 MΩ | 0.035 + 0.005 | 0.001 + 0.0005 || * Add 15 µV if not immediately following offset null. ---- === DC Current Accuracy === |<100% 25%>| ^ **Range** | **Burden Voltage** | **1-Year Accuracy ±(% of Reading + % of Range)** | ** Temperature Coefficient ±(% of Reading + % of Range)/C° ** || ^ 10 mA | <0.03 V | 0.070 + 0020 | 0.0035 + 0.0010 || ^ 100 mA | <0.3 V | 0.070 + 0.003 | 0.0020 + 0.0010 || ^ 1 A | <0.03 V | 0.130 +0.025 | 0.0065 + 0.0010 || ^ 10 A* | <0.3 V | 0.130 + 0.004 | 0.0045 + 0.0010 || * 30 seconds on, 30 seconds off. Add 300 ppm for currents >2.2 A. After measuring >5 A, Wait two minutes to get full accuracy in the 1 A range. ---- === DC Resistance Accuracy (2-Wire)*, 1V Open Circuit Voltage === |<100% 25%>| ^ **Range** | **Short-Circuit Current** | **1-Year Accuracy ±(% of Reading + % of Range) ** | ** Temperature Coefficient ±(% of Reading + % of Range)/C° ** || ^ 100 Ω | 170 µA | 0.018 + 0.050 | 0.0010 + 0.0005 || ^ 1 kΩ | 170 µA | 0.018 + 0.005 | 0.0010 + 0.0005 || ^ 10 kΩ | 70 µA | 0.018 + 0.005 | 0.0010 + 0.0005 || ^ 100 kΩ | 10 µA | 0.018 + 0.005 | 0.0010 + 0.0005 || ^ 1 MΩ | 1.1 µA | 0.035 + 0.005 | 0.0040 + 0.0005 || ^ 10 MΩ | 1.1 µA | 0.150 + 0.005 | 0.0100 + 0.0005 || ^ 100 MΩ | 1.1 µA | 1.3 + 0.005 | 0.1000 + 0.0005 || * Perform offset nulling ---- **Caution:** The input terminals of the DMM are not protected for electromagnetic interference. As a result, the DMM may experience reduced measurement accuracy or other temporary performance degradation when connected to unshielded test leads in an environment with radiated or conducted radio frequency electromagnetic interference. |<100% 25%>| ^ DC Continuity Accuracy Range | 100 Ω || ^ DC Diode Test Range | 2 V || ^ Effective Common-Mode Rejection-Mode Ratio (CMRR), 1 kΩ resistance in LO lead | >100 dB || ^ Normal-Mode Rejection Ratio (NMRR), 50/60 Hz ± 0.1% | >100 dB || ^ Overrange | 105% of range except 300 V || ---- ====AC==== ===AC Voltage Accuracy=== |<100% 25%>| ^ **Range (rms)** | **Peak Voltage**| **Frequency** | ** 1-Year Accuracy ±(% of Reading + % of Range)**|**Temperature Coefficient ±(% of Reading + % of Range)/C°**|| ^ 100 mV, 1 V, 10 V, 100 V, 265 V | ±210 mV, ±2.1 V, ±21 V, ±210 V, ±400 V | 20 Hz to 45 Hz | 0.91 + 0.10 | 0.01 + 0.005 || ^ ::: | ::: | 45 Hz to 65 Hz | 0.30 + 0.05 | 0.01 + 0.005 || ^ ::: | ::: | 65 Hz to 1 kHz | 0.21 + 0.05 | 0.01 + 0.005 || ^ ::: | ::: | 1 kHz to 5 kHz | 0.12 + 0.05 | 0.01 + 0.005 || ^ ::: | ::: | 5 kHz to 20 kHz | 0.35 + 0.05 | 0.01 + 0.005 || ---- ===AC Current Accuracy=== |<100% 25%>| ^ **Range**|** Short-Circuit Current**| **Burden Voltage (rms)**| **Frequency**| ** 1-Year Accuracy ±(% of Reading + % of Range)** | ** Temperature Coefficient ±(% of Reading + % of Range)/C° ** || ^ 5 mA | ±10.5 mA | <0.02 | 20 Hz to 1 kHz | 0.20 + 0.01 | 0.01 + 0.005 || ^ 5 mA | ±10.5 mA | <0.02 | 1 kHz to 5 kHz | 0.60 + 0.01 | 0.01 + 0.005 || ^ 50 mA | ±105 mA | <0.2 V | 20 Hz to 1 kHz | 0.20 + 0.01 | 0.01 + 0.005 || ^ 50 mA | ±105 mA | <0.2 V | 1 kHz to 5 kHz | 0.50 + 0.01 | 0.01 + 0.005 || ^ 500 mA | ±1.05 A | <0.02 V | 20 Hz to 1 kHz | 0.15 + 0.01 | 0.01 + 0.005 || ^ 500 mA | ±1.05 A | <0.02 V | 1 kHz to 5 kHz | 0.50 + 0.01 | 0.01 + 0.005 || ^ 5 A | ±1.05 A | <0.02 V | 20 Hz to 1 kHz | 0.25 + 0.03 | 0.01 + 0.005 || ^ 5 A | ±1.05 A | <0.02 V | 1 kHz to 5 kHz | 0.60 + 0.03 | 0.01 + 0.005 || ---- ===== DC Power Supply ===== ^ Outputs | 0 V to +6 V | 0 A to 1 A || ^ ::: | 0 V to +25 V | 0 mA to 500 mA (isolated) || ^ ::: | 0 V to -25 V | 0 mA to 500 mA (isolated) || **Note:** the +25 V and –25 V channels are bank isolated from ground but not from each other. ---- === DC Accuracy / Resolution === |<100% 25%>| ^ **Output** | ** Type ** | ** +6 V **| **+25 V**|**-25 V**|| ^ DC Output | Voltage | 0 V to +6 V | 0 V to +25 V | 0 V to -25 V || ^ ::: | Current | 1 A | 500 mA | 500 mA || ^ Programming Accuracy, ±(% of reading + offset) | Voltage | 0.1% + 5 mV | 0.15% + 20 mV | 0.15% + 20 mV || ^ ::: | Current | 0.2% + 10 mA | 0.15% + 4 mA | 0.15 + 4 mA || ^ Readback Accuracy, ±(% of reading + offset) | Voltage | 0.1% + 5 mV | 0.15% + 20 mV | 0.15% + 20 mV || ^ ::: | Current | 0.2% + 10 mA | 0.15% + 4 mA | 0.15% + 4 mA || ^ Programming Resolution | Voltage | 1.7 mV | 6.5 mV | 6.5 mV || ^ ::: | Current | 0.30 mA | 0.15 mA | 0.15 mA || ^ Readback Resolution | Voltage | 0.41 mV | 1.7 mV | 1.7 mV || ^ ::: | Current | 70 µA | 35 µA | 35 µA || ^ Load Regulation ±(% of reading + offset) | Voltage | 0.01% + 25 mV | 0.03% + 5 mV | 0.03% + 5 mV || |<100% 25%>| ^ Overvoltage Protection | 30 V (25 V channels) and 10 V (6 V channel) || ^ Reverse Voltage Protection | Reverse clamp diode, protected by self-resetting fuse || ---- ===== External Trigger (TRIG) ===== |<100% 25%>| ^ Direction Control | Input or output, software-selectable || ^ Logic Level | 5 V compatible TTL input, 3.3 V LVLTTL output || ^ Drive Strength | 4 mA || ^ Input Voltage | 0 V to 5 V || **Note:** The external trigger line is designed to withstand accidental overvoltage from signals on the ADP5250 or similar devices. It is not recommended for use with signals likely to exceed 0 V to 5 V in normal operation. ---- ===== Features ===== ==== Spectrum Analyzer ==== |<100% 25%>| ^ Frequency Range | DC to maximum bandwidth of scope || ^ Display Modes |Magnitude, average, peak hold, min hold, count || ^ Y Axis | Logarithmic (dbV, dBu, dBm) or linear (volts) || ^ X Axis | Linear or logarithmic || ^ Windowing Functions | Rectangular, Triangular, Hamming, Hann, Cosine, Blackman-Harris, Flat Top, Kaiser || ---- ==== Network Analyzer ==== |<100% 25%>| ^ Frequency Range | 1 mHz to 10 MHz, up to 10k samples per decade || ^ Display Modes | Magnitude, Phase || ^ Y Axis | Linear or logarithmic || ^ X Axis | Linear or logarithmic || ^ Plots | Time, FFT, Nichols, Nyquist || ---- ==== Math Channels ==== |<100% 25%>| ^ Operations | addition “+”, subtraction “-”, multiplication “*”, division “/”, reminder “%”. || ^ Brackets | parentheses (), square brackets [] || ^ Constants | Exp, Ln, Log, Pi || ^ Functions | Logarithm, power, minimum, maximum, square root, sine, cos, tan, arccos, arctan, arctan2, absolute value, round, floor, ceil || ^ Operands | All analog and digital input channels, reference waveforms, time, constants, Pi || ---- ==== Automatic Measurements (Analog Channels Only) ==== |<100% 25%>| ^ Oscilloscope Mode | DC to maximum bandwidth of scope || ^ Spectrum Mode | Magnitude, average, peak hold, min hold, count || ^ Statistics | Logarithmic (dbV, dBu, dBm,) or linear (volts) || ---- ===== Connectivity ===== ==== Wired USB Interface ==== |<100% 25%>| ^ USB Specification | USB 2.0 Hi-Speed || ---- ===== Power Requirements ===== |<100% 25%>| ^ Voltage Input Range | 100 V AC to 240 V AC, ±10%, 50/60 Hz || ^ Power Consumption | 100 W maximum || ^ Power Input Connector | IEC C13 power connector || ^ Power Disconnect | The AC power cable provides main power disconnect. Do not position the equipment so that it is difficult to disconnect the power cable. Depressing the front panel power button does not inhibit the internal power supply. || ---- ===== Physical Characteristics ===== |<100% 25%>| ^ Enclosure | 25.40 cm x 19.05 cm x 7.77 cm \\ (10.00 in x 7.50 in x 3.06 in) || ^ Weight | 2.05 kg (4 lb 8.3 oz) || ^ Security Cable Slot | 1, complies with Kensington security slot dimensions || ---- **Note:** If the device needs to be cleaned, wipe with a dry towel. **Note:** Use the ADP5250 in a horizontal orientation. Allow at least 10.16 cm (4.0 in.) of clearance in front and behind the ADP5250 for USB, power, and common connector cabling. ===== Safety Voltages ===== Connect only voltages that are within these limits. ==== DMM Isolation Voltages (Ground to Earth) ==== {{:test-and-measurement:analog-discovery-pro-5250:hazard.png?40}} **Hazardous Voltage:** This icon denotes a warning advising you to take precautions to avoid electrical shock. === Channel-to-earth Ground === |<100% 25%>| ^ Continuous | 300 V, Measurement Category II || ^ Withstand | 3000 V (RMS), verified by a 5 s dielectric withstand test || ---- Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe. {{:test-and-measurement:analog-discovery-pro-5250:caution.png?40}} **Caution:** Do not connect the ADP5250 hardware to signals or use for measurements within Measurement Categories III or IV. ==== DC Power Supply Isolation Voltages ==== |<100% 25%>| ^ +25V and -25V to earth ground, continuous | 60 VDC, Measurement Category 1 ||| **Note:** //Measurement Categories CAT I and CAT O (Other) are equivalent. The input circuits are not intended for direct connection to the MAINs building installations of Categories CAT II, CAT III, or CAT IV.// ---- =====Environmental===== |<100% 25%>| ^ Operating Temperature | 0 °C to 40 °C || ^ Storage Temperature | -20 °C to 60 °C || ^ Operating Humidity | 10% to 90% RH, non-condensing \\ DMM full accuracy at 10% to 80% || ^ Storage Humidity | 5% to 95% RH, non-condensing || ^ Cooling | Forced air circulation (positive pressurization) through a fan. Fan speed automatically adjusts according to operating conditions. Intake and exhaust locations are on rear of device. Ensure that the intake and exhaust locations are not obstructed. || ^ Pollution Degree | 2 || ^ Maximum Altitude | 2,000 m || **Note:** For indoor use only. ---- ==== Shock and Vibration ==== |<100% 25%>| ^ Operational Shock | 30 g peak, half sine, 11 ns pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) || ^ Random Vibration (Operating) | 5 Hz to 500 Hz, 0.3 grms || ^ Random Vibration (Nonoperating) | 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) || ---- {{page>test-and-measurement:analog-discovery-pro-5250:certifications&fullpage&noindent}} ---- All specifications provided courtesy of NI's documentation for the VB-8012: https://www.ni.com/docs/en-US/bundle/vb-8012-specs/resource/371527e.pdf.