====== OpenScope MZ - Calibration ====== {{ youtube>Bxf24fgrgKM?large }} The OpenScope MZ uses software calibration to improve measurement accuracy and each OpenScope MZ is calibrated as part of the manufacturing process. The OpenScope MZ **does not** need to be re-calibrated before each use, however, calibration is recommended before the first use and when changing environments, especially if there is a large temperature variation between environments. Calibration tables are saved in the OpenScope MZ flash in one of two locations: **flash** and **SD**. The **SD** calibration is intended to be overwritten by the user during normal use. The **flash** calibration is populated during manufacturing but it can also be overwritten. Calibration data is applied to the instruments when the OpenScope MZ boots. If the **SD** calibration exists it will be applied. If not and the **flash** calibration exists it will be applied. If neither the **SD** nor **flash** calibrations exist the ideal calibration will be applied to the instruments. ==== Calibration Process ==== The OpenScope MZ is calibrated using WaveForms Live. - Launch [[http://waveformslive.com/ | WaveForms Live]]. - [[software:waveforms-live:add-device| Add the OpenScope MZ]] to WaveForms Live. \\ - Click the **More** button on the device card and select **Configure**. \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-4.png |}} \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-5.png |}} \\ \\ - Under the **Calibration** section click **Calibrate**. \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-6.png |}} \\ \\ - Device specific instructions are provided. - Connect the solid red wire to the solid orange wire. - Connect the solid white wire to the solid blue wire. - Click **Next** to begin calibration. \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-7.png |}} \\ \\ - The calibration process takes about 30 seconds. \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-8.png |}} \\ \\ - The calibration information is displayed. Select the desired **storage location** (flash is recommended) and click **Save**. Then click **Done**. \\ \\ {{ reference:test-and-measurement:openscope-mz:waveforms-live-openscope-calibration-9.png |}} \\ \\ - The new calibration data has been applied to the instrument and will be applied at boot. {{tag>instrumentation tutorial openscope-mz calibration}} {{include_n>4}}