====== OpenLogger - Calibration ======
===== First-Time Calibration =====
Before being used for the first time, a new OpenLogger must first be calibrated. This process determines constants used to accurately calculate analog voltages for data points being logged, and stores these constants in non-volatile memory.
**Note:** //After first-time calibration, described in this guide, a device can be recalibrated, if desired, through its Configuration Menu, found through the Device Manager in WaveForms Live.//
----
=== 1. ===
When the device is connected to WaveForms Live for the first time, a prompt to calibrate the OpenLogger will appear. Click **OK**.
{{ reference:test-and-measurement:openlogger:uncalibrated_device_prompt.png?500 |Calibrate your device}}
=== 2. ===
Connect the provided 2x18 flywires to the OpenLogger with the **__nub side facing towards the center__** of the OpenLogger.
----
=== 3. ===
Connect the red wire on pin 6 (DC OUTPUT 1) to the solid orange wire on pin 35 (Analog Input 1). Take a look at the [[test-and-measurement:openlogger:reference-manual#openlogger_pinout|OpenLogger Pinout]] in the Reference Manual for details on which wire corresponds to which OpenLogger pin.
{{ /learn/instrumentation/tutorials/openlogger/calib-1.png?500 |Connect the red and orange wires}}
=== 4. ===
In WaveForms Live, click the **Begin** button to begin calibrating Analog Input 1.
{{ reference:test-and-measurement:openlogger:calibration_step_1.png?500 |Begin the calibration process}}
=== 5. ===
Wait for the calibration initialization to complete.
{{ reference:test-and-measurement:openlogger:starting_calibration_process.png?500 |Calibration progress bar}}
=== 6. ===
Connect the solid blue pin 33 (Analog Input 2 to the same red wire previously used and click the **Next** button.
=== 7. ===
After calibrating Analog Input 2, continue through the remaining 6 analog inputs (the pink, green, gray, purple, yellow, and brown wires) until calibration is complete. Once all analog inputs have been calibrated, click the **DONE** button.
{{ /learn/instrumentation/tutorials/openlogger/calib-2.png?500 |Start step 2}}
=== 8. ===
A final calibration complete screen will appear. Click the **DONE** button - see the image to the right.
----
=== 9. ===
Click the **DONE** button at the bottom of the page to back out to the configuration screen.
----
=== 10. ===
In the device configuration screen, click the back arrow at the top left of the page to return to the device selection screen.
{{ reference:test-and-measurement:openlogger:calibration_complete.png?500 |Calibration complete}}
----
===== More Information =====
Check out the OpenLogger's [[test-and-measurement:openlogger:start|Resource Center]] for more guides, tutorials, and projects.